- End cicle: When sterilization is finished, cold process starts. In case
of solids sterilization, fast steam output can be made using the
STEAM
button. The unsteaming is made by the output (11).
YOU CAN NOT OPEN THE LID UNTIL THE PRESSURE IS NOT
0 BAR AND THE INTERNAL TEMPERATURE UNDER 95ºC. SEE
SECTION 8.1
DO NOT OPEN THE VALVE UNTIL THE CYCLE IS
COMPLETELY FINISHED TO AVOID POSSIBLE
BURNS WITH HEAT WATER OR STEAM.
¡WARNING!
- THE DIRECT CONTACT WITH PRESSURE
STEAM CAN PRODUCE SERIOUS BURNS.
- IF YOU USE A HOSE TO MAKE THE STEAM
OUTPUT, DO NOT HOLD IT WITH HANDS TO
AVOID BURNS.
9 Settings
From the home screen, press the
key to go to
Settings
.
-Language: allows you to select the language of messages
-Brightness: adjusts the screen brightness
-Printer: Yes/No
-Date/Time: sets the date and time of the system
-Insert password:
0000
allows adjusting the temperature offset.
Note: Correction range from -9.9 to +9.9ºC. The entered va-
lue is added or substracted to the measured temperature by
correcting it.
-Download logs: transfers data stored in memory to a USB pen
-Units: Changes the numeric format
-Information: software version and version update button
By pressing the CAL button, you enter Calibration mode and allow you
to adjust the temperature offset.
When the lid is ready to be opened, the display (6) shows the word
CORRECT CYCLE
.
If liquids or culture media are sterilized must leave cooler until the
pressure will be 0 bar. During the cooling, is the same case of manual
depressurize, the display shows the internal temperature.