8.3 Program selection
- To select the program, use the arrow keys
< >
8.4 Program modification
- Press on the temperature and / or the program time and modify its
value with the keys
+ -
.
These values are automatically saved.
8.5 Cycle starting
- Once the temperature and time have been adjusted, press the RUN
key to start the process.
8.6 Work cycle steps
1 Purging: The autoclave expels the air inside the boiler to obtain
saturated steam. The
Presoclave III Plus
autoclaves perform
what we call an atmospheric purge. The temperature increases
until reaching 103°C. The purge solenoid valve remains open for 3
minutes once it reaches 103°C.
2 Temperature rise to the setpoint temperature: The temperature
continues risING from 103ºC to the setpoint temperature, with all
valves closed.
3 Sterilization: 20 seconds after the setpoint temperature is reached,
the autoclave begins the sterilization that will continue throughout
the selected set time. The equipment purges cyclically to ensure
proper operation.
4 Cooling: At the end of the sterilization, the equipment stops heating
and the cooling phase begins.
NOTE: when the
STEAM
key appears, if pressed, the depressurization
valve opens. This function is useful when you want to accelerate the
cooling process or to ensure that there is no pressure in the chamber.
BEFORE STARTING THE CICLE, ALWAYS
CHECK IF THERE IS ENOUGH WATER IN THE
CHAMBER. IF NOT, THE HEATING ELEMENT
COULD BE DAMAGED.