59
2) User Calibration of Measured Value
1. Touch the
[User Calibration]
in the Hardware
Setting screen to display the User Calibration
screen.
2. Using the button of
[+0.1EV]
or
[-0.1EV]
at
bottom of screen, adjust calibration value in
0.1 step increments.
* In this screen, it is possible to take a
measurement and display the value, then
adjust by using the +/-0.1EV buttons. This
will be helpful if you would like to match the
measured value to another meter by reading
the same light source.
3. Touch the
[Close]
to complete the setting
and return to the Measuring screen.
3) Adjust Touch Panel Display Position
1. Touch the
[Adjust Touch Panel]
in the
Hardware Setting screen to display the Touch
Panel Adjusting screen.
2. A white cross mark will appear at the upper
left of the screen. Please touch this cross.
A red cross mark will be displayed in the
touched position.
3. The second white cross mark will appear
in another position. Touch it with finger to
display a red cross mark in the touched
position. Repeat this touch-and-display
process eight times in total.
4. Touch the
[OK]
in the screen saying [Data
is determined by pressing the “OK”.]
to complete adjustments and return to
Measuring screen.
Touch the
[Cancel]
to cancel adjustments.
User Calibration Screen
Calibration
value
Touch Panel Adjusting Screen
Touch Panel Adjust
Confirmation Screen
White
cross
(cursor)
Touched
position