Potential Redox Calibration Menu (O.R.P.)
Pressing the CAL key for 3 seconds takes you into the calibration menu. If calibration was excluded during
programming, the following appears on the display:
Calibration
Off
If calibration is active:
Calibration
Automatic
Calibration
Automatic
Automatic Cal.
Wait.... 60s
Buffer Sol. 465mV
Buffer Sol. 465mV
Quality 100%
|||||| ok 465mV
Automatic Cal.
Error 465mV
Cal
-
Calibration
Manual
Automatic Cal.
Wait.... 60s
Buffer Sol. 465mV
Buffer Sol. 600mV
Quality 100%
|||||| ok 7 pH
Automatic Cal.
Error 7 pH
+
-
It is possible to select automatic or manual mode.
-
Automatic calibration:
The buffer solution value appears on the display. Insert the probe in the bottle and press the
key. A 60
second countdown necessary to complete calibration will appear on the display. If the alignment quality is below
50%, an error message appears on the display and you should press
to exit calibration (the pump exits
automatically after 4 seconds). If the quality is above 50%, the value is shown on the display and you should press
the
key to complete the procedure.
-
Manual calibration:
The buffer solution value appears on the display. Insert the probe in the bottle and press the
key. The value
of 465 mV should now flash on the display. Insert the probe in your solution and use the
keys to display
the value of the solution in your possession, then confirm by pressing the
key and begin the calibration
procedure as before