Seko TEKNA TPR 600 User Manual Download Page 10

Alarms 

Display Cause 

Interruption 

Fixed alarm LED 
Flashing word “Lev” 
I.e.  
 

Man           
Lev           P100% 

 

 

 

 

End of level alarm, without 
interrupting pump operation 

Restore the liquid level. 

Fixed alarm LED 
Flashing words “Lev” and “stop” 
I.e.  
 

Man           
Lev Stop      P100% 

 

 

 

 

End of level alarm, with interruption 
to pump operation 

Restore the liquid level. 

Flashing word “Mem” 
I.e.  
 

1:n              6           
Mem            

 

 

 

 

The pump receives one or more 
impulses during dosage with 
memory function on Off 

Press the 

 key 

Flashing word “Mem” 
I.e.  
 

1:n       

M

      6           

Mem            

 

 

 

 

The pump receives one or more 
impulses during dosage with 
memory function on On 

When the pump finishes receiving 
external impulses, it returns the 
memorised strokes  

Fixed alarm LED 
Flashing word “Flw” 
I.e.  
 

Man    

F

       

Flw           P100% 

 

 

 

 

Active flow alarm. The pump has 
not received the programmed 
number of signals from the flow 
sensor. 

Press the 

 key 

 
I.e.  
 

Parameter Error    PROG 
to default 

 

 

 

 

Communication error with the 
eeprom. 

Press the 

 key to restore the 

default parameters. 

Flashing word “OFA” 
Flashing word “stop” 
I.e.  
 

High        475 mV OFA 
Stop     P 75% 

 

 

 

 

O.F.A. alarm 

Press the 

 key to stop the 

flashing word “stop”. Press the key 
again to start up the pump again. 

Flashing word “Alm” 
I.e.  
 

High        475 mV Alm        
P 75% 

 

 

 

 

The probe reading is outside the set 
alarm band range 

Make sure that the “Alarm Band” 
parameter is set correctly in the 
programme 

Flashing word “Cal” 
I.e.  
 

High        475 mV Cal        
P 75% 

 

 

 

 

Probe not calibrated alarm 

Calibrate the probe 

 

 

EM00136063 rev. 

1.0 

10

Summary of Contents for TEKNA TPR 600

Page 1: ...nal on the display Used to exit the various menu levels Before definitively exiting the programming phase you will be asked if you wish to save any changes Access to the pump calibration menu If in Off mode the calibration menu is not activated Used to run upwards through the menu or increase the numerical values to be changed Can be used to start dosage in Batch mode Used to run downwards through...

Page 2: ...ous programming levels Upon exiting programming the display will show Exit Exit No Save Save to confirm the selection PROG Temperature Man 25 C Probe Calib On Configuration Pump Functions Max flow rate 100 Alarm Relay N Open Language English O F Alarm 0m Resolution 0 01 Alarm Band 2 00 pH Prop Band 0 50 pH Setpoint Type Acid Setpoint 7 00 pH Prog Configuration Pump Functions pH Measure Pump Functi...

Page 3: ...mode The flow can be manually regulated by pressing the keys at the same time to increase the flow or the keys to decrease it Display during Operation Display during Setting MODE key MAN F F F Lev Stop P100 Operating mode Man Manuel Alarms and statuses Liv Level alarm Fls Debit alarm FLOW sensor status Pump status Empty pump in start Stop pump stationary Paus pump in pause Current dosage value dep...

Page 4: ...l is triggered if the pH value does not reach the set point The measurement resolution 1 or 2 decimal points Deactivation activation of the calibration procedure Manual temperature value in ºC default or ºF The maximum frequency can be modified during operation by pressing the keys at the same time to increase the flow or the keys to decrease it Display during Operation Display during Setting MODE...

Page 5: ... at the same time to increase the flow or the keys to decrease it Display during Operation Display during Setting MODE key High F F F 560mV Lev Stop P100 Setpoint type High Low Alarms and statuses Cal calibration not completed Lev Level alarm Flw Flow alarm Alm Measurement outside alarm band OFA O F A alarm Flow sensor status Potential Redox measurement value Pump status Empty pump in start Stop p...

Page 6: ...ike using the keys and confirm by pressing the key AUTOMATIC the pump makes 100 strikes which are started by pressing the key At the end of this process enter the quantity sucked up by the pump using the keys and confirm by pressing the key The entered figure will be used in flow calculations Paragraph 7 Statistics Programming Operation Statistic Statistic Statistic Statistic Reset Hours 10 Hours ...

Page 7: ...key and you can then use the keys to set the value Confirm by pressing the key Press to return to the main menu Paragraph 10 Level Alarm Programming Operation Alarm Level Alarm Flow Alarms Level Stop Alarms Level Alarm Stop Alarm PROG ESC This makes it possible to set the pump when the level sensor alarm is activated In other words you can decide whether to stop dosage Stop or simply activate the ...

Page 8: ...ss the key A 60 second countdown necessary to complete calibration will appear on the display If the alignment quality is below 50 an error message appears on the display and you should press to exit calibration the pump exits automatically after 4 seconds If the quality is above 50 the value is shown on the display and after pressing the key the buffer solution at pH 4 or 9 will be requested At t...

Page 9: ...ppears on the display Insert the probe in the bottle and press the key A 60 second countdown necessary to complete calibration will appear on the display If the alignment quality is below 50 an error message appears on the display and you should press to exit calibration the pump exits automatically after 4 seconds If the quality is above 50 the value is shown on the display and you should press t...

Page 10: ...xternal impulses it returns the memorised strokes Fixed alarm LED Flashing word Flw I e Man F Flw P100 Active flow alarm The pump has not received the programmed number of signals from the flow sensor Press the key I e Parameter Error PROG to default Communication error with the eeprom Press the key to restore the default parameters Flashing word OFA Flashing word stop I e High 475 mV OFA Stop P 7...

Reviews: