Paragraph 5 – Setting the Alarm Relay
Programming Operation
-
-
-
-
Pump Functions
Max flow rate
Alarm Relay
P100%
N.Open
PROG
+
-
This is used to set the alarm relay in the absence of an
alarm situation, if open (default) or closed.
Changes can be made by pressing the
key, then using
the
keys to set the new value. Press
to
confirm and return to the main menu
Paragraph 6 – Flow Calibration
Programming Operation
-
-
-
-
Pump Calibration
Pump Calibration
Automatic Cal.
Automatic Cal.
Automatic Cal.
Pump Calibration
0,23 cc/stroke
cc/stroke 0,23
Start 100 strok.
Strokes 100
ml 20
Manual
PROG
+
-
+
-
Pump Calibration
Automatic
The memorised cc value per strike appears in the main
menu. It can be calibrated in two different ways:
MANUAL – manually enter the cc value per strike using the
keys and confirm by pressing the
key
AUTOMATIC – the pump makes 100 strikes, which are
started by pressing the
key. At the end of this process,
enter the quantity sucked up by the pump using the
keys and confirm by pressing the
key.
The entered figure will be used in flow calculations.
Paragraph 7 - Statistics
Programming Operation
-
-
-
-
-
-
Statistic
Statistic
Statistic
Statistic Reset
Hours 10
Hours 10
Strokes 1000
NO
PROG
+
-
Statistic
Statistic
Statistic
Q.ty(L) 100
Power 10
Reset
ESC
The main menu displays the pump operation times. By
pressing the
key you can access other statistics:
- Strokes = number of strokes made by the pump
- Q.ty (L) = quantity dosed by the pump in litres; this figure is
calculated on the basis of the memorised cc/stroke value
- Power = number of pump starts
- Reset = use the
to reset the counters (YES) or
otherwise (NO), then confirm by pressing the
key.
Pressing the
key will take you back to the main menu.