1
0
SetUp
Default
Set-Up
Settings
Rinstrum R320
/323
Indicator
s
K356 v4.16
CAL
ZERO
-
(calibrate)
SPAN
- (cali
brate)
FAC.CAL
-
(Restores factory default settings for BUILD + CAL )
(Not used)
SERIAL
Refer to manual for Serial and PRINT setup
PRINT BAUD =
9600
BITS = 8
DTR = off
FORMAT = PRINT B
TEST
SPEC
(Do not use)
Tech Support 24-7
* * *
Call
307-321-4505
BUILD
SCALE = "0.00000" Number related to pressure
SAFE.PC - Default setting "000000".
Requires FULL PC passcode to be set up. FULL.PC
(IMPORTANT: If you forget this code, reset at factory is required!)
OPTION U S E = i
NDUST
FILTER =
0.5
MOTION =
0.5
-1.0
INIT.Z =
OFF
Z.TRAC =
OFF
Z.RANGE = -2_ 2
Z.BAND = 0000
0
0
R.RENTRY = OFF
DP = 000000 (no decimals)
CAP = 5000
RES
=
1
DUAL = 60000
UNITS = lb
HI.RES =
OFF
CABLE = 4
Navigation:
Press
ZERO
Button to select major setting Group
Press
TARE
Button to select Setting
Press
SELECT
Button to acivate selected Setting
Use
SELECT
and
Buttons to select and adjust
Settings
Use
f
/OK
Button to save the selected Setting.
f
/OK
Button saving is temporary. You MUST use
the
ZERO
Button to go to
END
, then press the
TARE
Button to secure and save the
Full SetUp
.
After
Saving
is displayed, the Indicator will power-up
in the Weighing Mode
and display
READY
.
All settings, once se
t
and
s
aved will remain in
memory until the Settings are changed in the Setup
mode.
Power - On/Off does not affect SetUp.
FACTORY - (Restores all factory default settings except Calibration) (Not used)
E
ND -
(
Press TARE and the
Display will
show
"SAVING" then Auto Power OFF to complete
Set-up process.)