![Seccua UrSpring Installation And User Instructions Manual Download Page 42](http://html1.mh-extra.com/html/seccua/urspring/urspring_installation-and-user-instructions-manual_1230352042.webp)
4
UrSpring Flow sensor (flow sensor K locking
ring (11354))
11351
5
UrSpring Power supply (cable + adapters)
10698
5. Troubleshooting
If your UrSpring unit reports an error, the red LED on the controller will blink.
5.1 Error codes
To view the error codes:
1. Go to menu
View alerts.
2. The display shows
Last Error ------
.
3. Use
and
to browse through the errors.
rror
Code
Possible cause
Solution
101
The emergency flush is running or
has run.
A flush is activated fo
ff r 20
seconds. Because the maximum
transmembrane pressure (feed to
filtrate) is exceeded (> 2.5
bar/
rr 36psi).
Seek and eliminate the cause of the high
transmembrane or absolute pressure. It can
damage the UrSpring
● Check the feed pressure and upstream
pressure reducing valve.
● Check if the pressure conditions or the
max. consumption has changed.
● Check if the filter module is fo
ff uled and
needs to be replaced.
102
Permeability of the filter module is
too low.
● Perf
rr o
ff rm a manual flush to verify
ff that the
flush is working.
● Check if the flush settings fit the
application and if they need optimization.
● Check if the filter module is fo
ff uled and
needs to be replaced.
Contact your serv
rr ice technician fo
ff r further help
Page 42/50
Summary of Contents for UrSpring
Page 1: ...Installation and User Instructions Seccua UrSpring April 2020 Version 1 10 Page 1 50...
Page 6: ...1 3 Overview of the UrSpring Page 6 50...
Page 53: ......
Page 54: ......
Page 55: ......
Page 81: ...Sie k nnen die Einweg Handschuhe nun ausziehen Seite 28...
Page 103: ...1 3 bersicht ber die UrSpring Seite 6...
Page 108: ...Installations und Benutzeranweisungen Seccua UrSpring Juni 2020 Version 1 10 Seite 1...