6.4 Compare reference spectrum files
Compare the change between two reference spectrum files. The amount of change is
related to the time interval between the updates and the amount of lamp use during that
time.
Note that this procedure is done automatically by the software when the user updates the
reference spectrum for the sensor. Refer to
on page 22 for
more information.
1.
From the Data menu, select
SUNA
, then
Compare Calibration
.
2.
Push
Browse
to find the first calibration, or reference file, to compare.
3.
Push
Browse
to find the second file to compare. Note that the files must be from the
same sensor.
4.
Push
Compare
.
A typical update interval of 3–6 months with no more than 100 hours of lamp use
should cause a change of no more than 10% in the 215–240 nm interval.
Below 215 nm, larger relative changes are normal.
Above 240 nm, the change is smaller than at the 215–240 nm range.
If there is a large change, do several reference updates 12–24 hours apart to monitor
the stability of the reference spectrum.
6.5 Update SUNA firmware
At regular intervals, make sure that the current firmware is installed in the sensor.
1.
Go to the manufacturer's web site to get the current firmware for the sensor.
2.
Save the firmware to the PC.
The firmware is an ".sfw" file.
3.
Make sure that the sensor is connected to the PC and a power supply.
4.
Push
Connect
.
5.
Go to the
Sensor
menu and select
SUNA
, then
Advanced
, then
Upload Firmware
File
.
Maintenance
25
Summary of Contents for SUNA V2
Page 2: ......
Page 6: ...SUNA quick start guide 4 ...
Page 14: ...Set up sensor and verify operation 12 ...
Page 16: ...Deployment and recovery 14 ...
Page 22: ...Transmit data 20 ...
Page 42: ...Reference 40 ...
Page 48: ...Software reference 46 ...
Page 52: ...General information 50 ...
Page 53: ......