SetTempA1=F
F=pressure temperature a1
SetTempA2=F
F=pressure temperature a2
SetPTCA0=F
F=pressure temperature compensation ptca0
SetPTCA1=F
F=pressure temperature compensation ptca1
SetPTCA2=F
F=pressure temperature compensation ptca2
PTCB0=F
F=pressure temperature compensation ptcb0
PTCB1=F
F=pressure temperature compensation ptcb1
PTCB2=F
F=pressure temperature compensation ptcb2
SetPOffset=F
F=pressure sensor offset correction, decibars
SetPFullScale=F
F=pressure sensor full scale, psia (set by manufacturer)
8.11 Hardware configuration
Hardware configuration commands are character strings up to 11 characters long. These
strings can include:
•
+
•
-
•
= (This cannot be the first character in the string.)
Temperature and conductivity sensor
SetTempSer=x
SetCondSer=x
The commands below are used to set auxiliary channel sensor types and serial number.
The sensor "type" is informational only and does not change the operation of the 25plus.
Auxiliary voltage sensor
SetVoltType0=
Set the serial number of the sensor installed on voltage channel 0. Same for each other Type 1-7.
SetVoltSN0=
Set the sensor name or model installed on voltage channel 0. Same for each other Type 1-7.
Auxiliary serial sensor
SetSerialTypeSer1=
Set the serial number of the sensor installed on serial sensor 1.
SetSerialSNSer1=
Set the sensor name or model installed on serial sensor 1.
SetSerialTypeSer2=
Set the serial number of the sensor installed on serial sensor 2.
SetSerialSNSer2=
Set the sensor name or model installed on serial sensor 2.
Reference: command descriptions
52
Summary of Contents for SBE 25plus
Page 2: ......
Page 6: ...SBE 25plus quick start guide 4 ...
Page 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Page 14: ...Figure 8 SBE 55 Overview 12 ...
Page 22: ...Set up sensor and verify operation 20 ...
Page 60: ...Serial sensor integration 58 ...
Page 64: ...General information 62 ...
Page 65: ......