b63–b32
cccccccc
conductivity, Hz
IEEE 754 format floating point, 32 bit
b95–b64
00pppppp
pressure
raw ADC counts*
b127–b96
00pppppp
pressure
temperature
raw ADC counts*
b143–b128
vvvv
voltage input 0
V
0
= (code ÷ 2
16
) × 5.000
b159–b144
vvvv
voltage input 1
V
1
= (code ÷ 2
16
) × 5.000
b175–b160
vvvv
voltage input 2
V
2
= (code ÷ 2
16
) × 5.000
b191–b176
vvvv
voltage input 3
V
3
= (code ÷ 2
16
) × 5.000
b207–b192
vvvv
voltage input 4
V
4
= (code ÷ 2
16
) × 5.000
b223–b208
vvvv
voltage input 5
V
5
= (code ÷ 2
16
) × 5.000
b239–b224
vvvv
voltage input 6
V
6
= (code ÷ 2
16
) × 5.000
b255–b240
vvvv
voltage input 7
V
7
= (code ÷ 2
16
) × 5.000
b287–b256
iiiiiiii
diagnostic
* Pressure and pressure temperature output does not have zero padding (pppppp, not
00pppppp).
The IEEE floating point numbers are decoded below:
31
30
23
22
0
S
Exponent (E)
Mantissa
Num = (-1)
S
× 2
(E-127)
× 1. M
Diagnostic and status information is not included in real-time data. It is stored in memory
and transmitted from there. The information below is saved with each scan:
Location
Hex
Variable
Encoding
b3–b0
000b
Vout fault
0=no fault, 1=fault
Bit position corresponds to channel
b7–b4
000b
Vout enable
0=no fault, 1=fault
Bit position corresponds to channel
b15–b8
0xii
auxiliary current
Raw ADC counts*
b23–b16
0xii
system current
Raw ADC counts*
b24
0b
memory full
File system always keeps a few 10s of kb at the end. This bit is
sent and the measurement is stopped.
b25
0b
battery low
Voltage input too low during deployment. This bit is sent and the
measurement is stopped.
b26
0b
serial overflow 1
Serial overflow occurred on this measurement on channel 1.
b27
0b
serial overflow 2
Serial overflow occurred on this measurement on channel 1.
b28
0b
pump enable
Pump is turned on.
b29
0b
generic error #1
Miscellaneous error
b30
0b
generic error #1
Miscellaneous error
b31
0b
generic error #1
Miscellaneous error
* To convert raw ADC counts to current (auxiliary or system, in mA) = 2.5 × raw ADC
counts ÷ 1024.
Deployment and recovery
26
Summary of Contents for SBE 25plus
Page 2: ......
Page 6: ...SBE 25plus quick start guide 4 ...
Page 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Page 14: ...Figure 8 SBE 55 Overview 12 ...
Page 22: ...Set up sensor and verify operation 20 ...
Page 60: ...Serial sensor integration 58 ...
Page 64: ...General information 62 ...
Page 65: ......