SetOutputformat=x
x=0: default. Format required for use with Seasave software. During data collection, transmit all T,
C, P data and selected auxiliary voltage sensor data in real-time, at the baud rate set by
SetBaudConsole=. Refer to the SetVOut commands below to select the voltage sensor channels
from which to see output.
x=1: Format required for use with autonomous water sample systems (SBE 32 with AFM or
SBE55). During data collection, set the baud rate to 9600, at 1 Hz. After data collection, the 25plus
goes back to the baud rate set by SetBaudConsole=
SetHistoricRate=x
Applicable only if the 25plus baud rate is set to 4800 and SetOutputFormat=0. The 4800 baud rate
is required when the 25plus is used with the SBE 33 or SBE 36 Deck Unit.
A slower transmission rate is required if the Deck Unit appends NMEA and surface PAR data.
x=0: The rate of output is 4 Hz if there are no external voltage sensors enabled for real-time output.
If there are 1 or more external voltage sensors enabled, set the rate of output to 2 Hz.
Use this setting if the SBE 33 or 36 append NMEA and surface PAR data.
x=1: The rate of output is 8 Hz if there are no external voltage sensors enabled for real-time output.
If there are 1 or more external voltage sensors enabled, set the rate of output to 4 Hz.
SetVOut0=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 0.
x=N: Do not show output.
SetVOut1=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 1.
x=N: Do not show output.
SetVOut2=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 2.
x=N: Do not show output.
SetVOut3=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 3.
x=N: Do not show output.
SetVOut4=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 4.
x=N: Do not show output.
SetVOut5=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 5.
x=N: Do not show output.
SetVOut6=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 6.
x=N: Do not show output.
SetVOut7=x
x=Y: if SetOutputFormat=0, show real-time data from voltage channel 7.
x=N: Do not show output.
Notes:
•
The 25plus always stores data from all 8 voltage channels in memory independently
from the settings for SetVOut0= through SetVOut7, and whether or not a sensor is
connected on each channel.
•
The 25plus configuration file (.xmlcon) must agree with the selection of external
voltages to be included with real-time data to see this data in Seasave and Data
Processing software. Look at and edit the configuration file in the software.
8.7 Serial sensor setup
When commanded to start data collection, the 25plus takes samples, stores the data in
memory, and transmits the data in real-time.
The first time the sensor is started after data collection (InitLogging or DeleteAll), all
previously stored data is overwritten. When the plunger switch is pulled out or Stop is
sent, the sensor stops data collection. Each time that data collection is started, the new
data is stored in a new directory. The data file has header information that shows the
incremented cast number, the date, time and sample numbers in that cast.
Reference: command descriptions
49
Summary of Contents for SBE 25plus
Page 2: ......
Page 6: ...SBE 25plus quick start guide 4 ...
Page 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Page 14: ...Figure 8 SBE 55 Overview 12 ...
Page 22: ...Set up sensor and verify operation 20 ...
Page 60: ...Serial sensor integration 58 ...
Page 64: ...General information 62 ...
Page 65: ......