TSVR
Supply power to all auxiliary voltage channels 0–7 and continuously measure and show raw counts. Push
Esc
to
stop and remove power.
TSV
Supply power to all auxiliary voltage channels 0–7 and continuously measure and show output voltages. Push
Esc
to stop and remove power.
GetVAux0
Supply power to voltage channels 0 and 1 (J0) and continuously measure and show output voltages.
Push
Esc
to stop and remove power.
GetVAux1
Supply power to voltage channels 2 and 3 (J1) and continuously measure and show output voltages.
Push
Esc
to stop and remove power.
GetVAux2
Supply power to voltage channels 4 and 5 (J2) and continuously measure and show output voltages.
Push
Esc
to stop and remove power.
GetVAux3
Supply power to voltage channels 6 and 7 (J3) and serial channels 1 and 2 (J6) and continuously
measure and show output voltages. Push
Esc
to stop and remove power. Serial data is not shown.
SetVAuxPower0=x x=Y: Supply power to voltage channels 0 and 1 to test.
x=N: Remove power
SetVAuxPower1=x x=Y: Supply power to voltage channels 2 and 3 to test.
x=N: Remove power
SetVAuxPower2=x x=Y: Supply power to voltage channels 4 and 5 to test.
x=N: Remove power
SetVAuxPower3=x x=Y: Supply power to voltage channels 6 and 7 and to serial channels 1 and 2 to test.
x=N: Remove power
Note: If power is supplied externally to a serial sensor, use SetVauxPower3=Y to supply
power to the connector to which the serial sensor is attached.
8.10 Calibration coefficients
F = floating point number. S = String with no spaces.
Calibration coefficients are set by the manufacturer and should be the same as the
Calibration Certificates that ship with the 25plus. They should also be the same as the
values stored in the .xmlcon configuration file of the 25plus.
F= floating point number. S = string with no spaces.
Temperature and conductivity calibration coefficients are not stored in the 25plus
EEPROM because these sensors are attached and can be changed in the field. These
coefficients are stored in the .xmlcon configuration file.
Auxiliary sensor calibration coefficients are not stored in the 25plus EEPROM.
•
Calibration coefficients for sensors that have raw data output are stored in
the .xmlcon file. Use the
Configure
menu in the Data Processing of Seasave
software to show or change these values.
•
Some serial sensors store calibration coefficients internally, and have data output that
show in engineering units. For sensors such as the SBE 38, connect the sensor
directly to a PC and use Seaterm, not Seaterm V2, to change the calibration
coefficients.
SetPCalDate=yyyy-mm-dd
Set pressure sensor calibration date and time (UTC)
SetPA0=F
F=temperature A0
SetPA1=F
F=temperature A1
SetPA2=F
F=temperature A2
SetTempA0=F
F=pressure temperature a0
Reference: command descriptions
51
Summary of Contents for SBE 25plus
Page 2: ......
Page 6: ...SBE 25plus quick start guide 4 ...
Page 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Page 14: ...Figure 8 SBE 55 Overview 12 ...
Page 22: ...Set up sensor and verify operation 20 ...
Page 60: ...Serial sensor integration 58 ...
Page 64: ...General information 62 ...
Page 65: ......