The software writes the data to the .xml or .txt file as applicable.
•
If "multiple files" was selected, the software does the sub-steps above for each
deployment.
When the data has been transmitted, the software shows the
S>
prompt.
10.
Verify that all of the data has been transmitted:
a.
Use Seasave to see raw hexidecimal data from the .xml file in engineering units.
b.
Use the SBE Data Processing software to process and make a plot of the data.
6.2 Transmit data with USB
Use the USB connection to transmit large quantities of data. It is much faster than RS232.
1.
Pull the plunger switch out to the Off position to stop the 25plus.
2.
Use a wrench if necessary to loosen and remove the battery end flange.
3.
Use a lint-free tissue or cloth to remove any water around the seams and O-ring
mating surfaces.
4.
Connect the manufacturer-supplied USB cable to the 25plus.
It may be necessary to try more than one cable because some pin lengths will not
connect correctly with the power and ground connectors.
5.
The connected PC opens a window with the contents and file structure of the 25plus
data by date.
6.
Copy or drag the files to a selected location on the PC.
7.
Make sure that all data has been transmitted:
a.
Use Seasave to show the raw hexidecimal data from the .xml file in engineering
units.
b.
Use SBE Data Processing software to process and make a plot of the data.
6.3 Convert data
Convert the .xml or .hex data to a .cnv (engineering units) file with the Data Conversion
module in the SBE Data Processing software. Make sure to verify that the .xmlcon
configuration file agrees with the 25plus before the data is converted.
1.
Start the Data Processing software.
2.
Select the Data Conversion module.
3.
Select the correct .xmlcon file as the
Instrument Configuration File
.
4.
Verify that the sensors and channels are correct and that the calibration coefficients
agree between the software and the sensor.
5.
Select the .xml or .hex data file as the "Input File."
6.
Specify the "Output Directory" for the converted file.
7.
Push
Start Process
. The converted data is saved to the specified "Output Directory."
8.
When the data is converted, use the other modules of the SBE Data Processing
software to align, filter, remove bad data, calculate derived variables, and make plots
of data.
Transmit and convert data
30
Summary of Contents for SBE 25plus
Page 2: ......
Page 6: ...SBE 25plus quick start guide 4 ...
Page 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Page 14: ...Figure 8 SBE 55 Overview 12 ...
Page 22: ...Set up sensor and verify operation 20 ...
Page 60: ...Serial sensor integration 58 ...
Page 64: ...General information 62 ...
Page 65: ......