20.
Examine the O-rings on the end flange. They must be pristine, with no lint or
scratches or chips.
•
Apply a small quantity of Parker Super O Lube
®
to any new O-rings.
21.
Align the end flange holes with the holes in the pressure housing.
22.
Carefully push the end flange into the pressure housing.
It may help to rotate the end flange so that the wires do not bend too much.
23.
If necessary, use a 9/64" hex key to remove the two screws from the end flange.
24.
Use a 9/64" hex key to install the two screws into the pressure housing of the sensor
again.
Set up sensor and verify function