Section 3: Setting Up System
SBE 45 Interface Box
15
Mode
Commands
IMode=x x=Normal
: On power-up, Interface Box is
in Normal mode for communicating with
Interface Box.
x=Connect45
: On power-up,
Interface Box is in SBE 45 mode, for
communicating with SBE 45 through
Interface Box. Equivalent to powering up
in Normal mode, and then immediately
sending
Connect45
.
When you send
IMode=Connect45
, Interface Box is also
immediately placed in SBE 45 mode.
Connect45
Put Interface Box in SBE 45 mode,
allowing you to send commands to
SBE 45 through Interface Box.
Connect38
Put Interface Box in SBE 38 mode,
allowing you to send commands to
SBE 38 through Interface Box.
ConnectNMEA
Put Interface Box in NMEA mode,
allowing you to look at data
as transmitted
by NMEA navigation device.
Example:
(user input shown in bold)
S>
CONNECT45
Connected to SBE 45, press @ or Esc to return to normal mode
S>
DS
(transmitted to SBE 45, so status response is from SBE 45)
SBE45 V 1.1 SERIAL NO. 0059
. . . .
(see SBE 45 manual for remainder of SBE 45 status response)
S>
OUTPUTSV=N
(do not output sound velocity from SBE 45)
@
mode switched back to normal
S>
DS
(transmitted to Interface Box, so status response is from Interface Box)
SBE45 Junction Box V 1.4
. . . .
(see above for remainder of Interface Box status response)
Example:
(user input shown in bold)
With the Interface Box in Normal mode, turn power on. Interface Box responds as follows:
S>SBE 45 Junction Box V 1.4
mode is normal
S>
IMODE=CONNECT45
(send command to change start-up mode; also switches to SBE 45 mode now)
Connected to SBE 45, press @ or Esc to return to normal mode
S>
DS
(transmitted to SBE 45, so status response is from SBE 45)
SBE45 V 1.1 SERIAL NO. 0059
. . . .
(see SBE 45 manual for remainder of SBE 45 status response)
S>
(send other commands to SBE 45 as desired)
@
(put Interface Box back in normal mode)
mode switched back to normal
S>
DS
(transmitted to Interface Box, so status response is from Box)
SBE45 Junction Box V 1.4
. . . .
(see above for remainder of Interface Box status response)
Turn power off and then on again. Interface Box starts up in SBE 45 mode:
S>SBE 45 Junction Box V 1.4
Mode is connect 45
SBE45 V 1.1
S>
(send commands to SBE 45 as desired)
Notes:
•
IMode=Connect45
is not
compatible with Seasave real-time
data acquisition software.
•
Type @ or press the Esc key to
exit SBE 45, SBE 38, or NMEA
mode and return to Normal mode
for communicating with the
Interface Box.
77
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...