Appendix III: Command Summary
SBE 45
45
Appendix III: Command Summary
CATEGORY
COMMAND DESCRIPTION
Status DS
Display status.
Baud=x
x=
baud rate (1200, 2400, 4800, 9600, 19200,
or 38400). Default 4800.
OutputFormat=x
x=0
(default): Output order is temperature,
conductivity, salinity, sound velocity.
x=1
: Suppress space before conductivity output.
x=2
: Reverse conductivity and salinity order (not for
use with Seasave, SBE Data Processing, or optional
Interface Box).
OutputCond=x
x=Y
(default): Output conductivity (S/m) with data.
x=N
: Do not.
OutputSal=x
x=Y
: Output salinity (psu) with data.
x=N
(default): Do not.
OutputSV=x
x=Y
: Output sound velocity (m/sec) with data,
calculated as defined by
SVAlgorithm=
.
x=N
(default): Do not.
SVAlgorithm=x
x=C
(default): Calculate sound velocity as Chen
and Millero.
x=W
: Calculate sound velocity as Wilson.
NCycles=x
x
= A/D cycles to average per sample (default=4).
Setup
QS
Quit session and place MicroTSG in quiescent (sleep)
state. Sampling stops. Applicable only if PCB J1
jumper in Normal position.
Interval=x
x=
interval (seconds) between samples (maximum
32767). Minimum time between samples determined
by
NCycles=
, desired parameters (salinity, etc.), and
baud rate.
AutoOff=x
(Functional only if J1 jumper in Normal position)
x=Y
: Go to sleep (enter quiescent state) if 2 minutes
have elapsed without receiving a command or without
sampling data.
x=N
: Do not automatically go to sleep.
J1 jumper - Normal
AutoRun=N
SingleSample=
Y
or
N
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, wait for a
command.
J1 jumper - Normal
AutoRun=Y
SingleSample=N
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, sample at
rate specified by
Interval=
. To stop sampling and get
S>
prompt, type
Stop
and press Enter key.
J1 jumper - Normal
AutoRun=Y
SingleSample=Y
Wake up when Connect on Toolbar clicked or Enter
key pressed while in quiescent (sleep) state, take and
output a single sample, automatically go to sleep. To
wake up and get
S>
prompt, type
Stop
and press
Enter key.
J1 jumper - Autopower
AutoRun=N
SingleSample=Y
or
N
Wake up when power applied, wait for a command.
J1 jumper - Autopower
AutoRun=Y
SingleSample=N
Wake up when power applied, sample at rate specified
by
Interval=
until power removed.
Required settings
for running MicroTSG in 3-wire (power, ground,
transmit) configuration or with Interface Box
.
J1 jumper - Autopower
AutoRun=Y
SingleSample=Y
Wake up when power applied, take and output a
single sample. Wait for another command until
power removed.
Go
Start sampling, as defined by
SingleSample=
and
Interval=
. Applicable if
AutoRun=N
, or
AutoRun=Y
and you previously sent
Stop
to
stop sampling.
Operating
PCB’s J1
jumper interacts
with these
commands:
•
Autopower –
pins 1 and 2
•
Normal –
pins 2 and 3
Stop
Stop sampling.
50
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...