2
To correct conductivity data taken between pre- and post-cruise calibrations:
islope = 1.0 + (b / n) [(1 / postslope) - 1.0]
where
islope = interpolated slope; this is the value to enter in the configuration (.con or .xmlcon) file
b = number of days between pre-cruise calibration and the cast to be corrected
n = number of days between pre- and post-cruise calibrations
postslope = slope from calibration sheet as calculated above (see
Appendix I:
Example Conductivity
Calibration Sheet
)
In the configuration (.con or .xmlcon) file, use the
pre-cruise calibration coefficients
and use
islope
for the value of
slope.*
Note:
In our SEASOFT V2 suite of programs, edit the CTD configuration (.con or .xmlcon) file using the Configure
Inputs menu in Seasave V7 (real-time data acquisition software) or the Configure menu in SBE Data Processing (data
processing software).
For typical conductivity drift rates (equivalent to -0.003 PSU/month), islope does not need to be recalculated more
frequently than at weekly intervals.
* You can also calculate preslope. If
α
is the conductivity computed from
post-cruise bath data
(temperature and
frequency) using
pre-cruise calibration coefficients
and
β
is the true conductivity in the
post-cruise bath
, then:
preslope =
(preslope is typically > 1.0)
In this case, pre-cruise calibration coefficients would be used and:
islope = 1.0 + (b / n) (preslope - 1.0)
Correcting for Conductivity Drift Based on Salinity Bottles Taken At Sea
For this situation, the
pre-cruise
calibration coefficients are used to compute conductivity and CTD salinity. Salinity
samples are obtained using water sampler bottles during CTD profiles, and the difference between CTD salinity and
bottle salinity is used to determine the drift in conductivity.
In using this method to correct conductivity, it is important to realize that differences between CTD salinity and
hydrographic bottle salinity are due to errors in conductivity, temperature, and pressure measurements, as well as
errors in obtaining and analyzing bottle salinity values. For typical Sea-Bird sensors that are calibrated regularly,
70 - 90% of the CTD salinity error is due to conductivity calibration drift, 10 - 30% is due to temperature calibration
drift, and 0 - 10% is due to pressure calibration drift. All CTD temperature and pressure errors and bottle errors must
first be corrected before attributing the remaining salinity difference as due to CTD conductivity error and proceeding
with conductivity corrections.
Σ
(
α
i
)(
β
i
)
n
Σ
(
α
i
)(
α
i
)
i=1
n
i=1
111
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...