Section 3: Setting Up System
SBE 45 Interface Box
16
Output Format
Command
Format=x
x=0
: Output standard format (default)
t1=ttt.tttt, c1=cc.ccccc, s =
sss.ssss, sv=vvvv.vvv, t2 =
ttt.tttt, lat = DD MM.MMMM W,
lon = DDD MM.MMMM W, hms =
HHMMSS, dmy = DDMMYY
x=1
: Output custom format for Kongsberg
EM 300 Multibeam Echo Sounder
tt.ttt vvvv.v
Sampling
Commands
Start
Start transmitting data.
Stop
Stop transmitting data.
Diagnostic
Command
*EETest
Test EEPROM; reset baud rates to defaults
(new bauds take effect on next power up).
Example:
(user input shown in bold)
S>
*eetest
wr: 128
. .
wr: 1920
rd: 256
. .
rd: 1792
passed EEprom test
S>
Note:
Interface Box starts transmitting data
automatically when power is turned on
and stops when power is removed.
Use
Stop
to stop transmitting data
temporarily while setting up the
instruments, and then restart
transmitting data by sending
Start
.
Note:
See
Data Output Format
in
Section 2:
Interface Box Description
for details.
•
Format=1
is not compatible with
Seasave real-time data acquisition
software or SBE Data Processing
post-processing software.
78
Summary of Contents for SBE 45 MicroTSG
Page 2: ...This page intentionally left blank ...
Page 4: ...Manual Generation Date 13 February 2012 1 ...
Page 6: ...This page intentionally left blank 3 ...
Page 7: ...4 ...
Page 112: ...109 ...
Page 118: ...6 Appendix I Example Conductivity Calibration Sheet 115 ...
Page 119: ...7 Appendix II Example Temperature Calibration Sheet 116 ...
Page 129: ...126 ...
Page 130: ...127 ...
Page 131: ...128 ...
Page 132: ...129 ...
Page 133: ...130 ...
Page 134: ...131 ...
Page 135: ...132 ...
Page 136: ...133 ...
Page 137: ...134 ...
Page 138: ...135 ...