SCS CTE701 Operating Instructions Manual Download Page 7

TB-9031

 Page 7 of 11

© 2022 DESCO INDUSTRIES INC

Employee Owned

SCS 

- 926 JR Industrial Drive, Sanford, NC 27332

East: (919) 718-0000 | West: (909) 627-9634 • Website: 

StaticControl.com

VERIFYING THE TOOL CIRCUIT

15. Disconnect the monitor's tool cord from its metal 

tool. 

16. Clip the red test lead's mini grabber to the tool cord. 

Figure 15. Connecting the Verification Tester to WS 

Aware Monitor's tool cord

Monitor's

Tool Cord

17. Wait for the monitor's tool LED to illuminate red and 

sound its audible alarm. 

18. Press and hold the Verification Tester's METAL 

GROUND PASS test switch. The monitor's tool LED 

will illuminate green, and its audible alarm will stop. 

This verifies the tool circuit's impedance limit. 

Figure 16. Pressing the METAL GROUND PASS test 

switch

19. Press and hold the Verification Tester's METAL 

GROUND FAIL test switch. The monitor's tool LED 

will illuminate red, and its audible alarm will sound. 

This verifies the tool circuit's impedance limit. 

20. Press and hold the Verification Tester's METAL 

GROUND PASS test switch. Simultaneously, press 

and hold the Verification Tester's EMI LOW test 

switch. The monitor's tool LED will remain green, 

and no audible alarm will sound. This verifies the 

operator circuit's low EMI voltage limit. 

Figure 17. Pressing the METAL GROUND FAIL test 

switch

Figure 18. Pressing the METAL GROUND PASS and 

EMI LOW test switches simultaneously

21. Press and hold the Verification Tester's METAL 

GROUND PASS test switch. Simultaneously, press 

and hold the Verification Tester's EMI HIGH test 

switch. The monitor's tool LED will blink red, and its 

audible alarm will sound. This verifies the operator 

circuit's high EMI voltage limit. 

Figure 19. Pressing the METAL GROUND PASS and 

EMI HIGH test switches simultaneously

22. Disconnect the red test lead from the monitor's tool 

cord. 

23. Reinstall the tool cord to the metal tool.

Summary of Contents for CTE701

Page 1: ...nds annual calibration of workstation monitors and the CTE701 Verification Tester The CTE701 Verification Tester meets ANSI ESD S20 20 and Compliance Verification ESD TR53 November 2022 The SCS CTE701 Verification Tester can be used with the following items Item Description 770067 WS Aware Monitor 770068 WS Aware Monitor CTC061 3 242 WW WS Aware Monitor CTC061 RT 242 WW WS Aware Monitor CTC062 RT ...

Page 2: ...operator circuit when pressed F Soft Ground Test Switch Simulates a MAT PASS condition on the monitor when pressed G Wrist Strap Test Switch Simulates a OPERATOR PASS condition on the monitor when pressed H Test Limit DIP Switch Configures the test limits on the CTE701 Verification Tester I High Metal Ground Test Switch Simulates a TOOL FAIL condition on the monitor when pressed J High EMI Test Sw...

Page 3: ...FF 14 ohms OFF ON OFF OFF 15 ohms ON OFF OFF OFF 16 ohms OFF OFF OFF OFF EMI The EMI high frequency signal is configured with switch 9 The CTE701 Verification Tester provides two different levels of high frequency signal elevated and normal Pressing the HIGH EMI pushbutton will load a high signal level within its range Pressing the LOW EMI pushbutton will load a low signal within its range Switch ...

Page 4: ...OLTAGE test switches simultaneously 6 Continue to press and hold the Verification Tester s WRIST STRAP test switch Simultaneously press and hold the Verification Tester s HIGH BODY VOLTAGE test switch The monitor s green operator LED will illuminate continuously its red LED will blink and an audible alarm will sound This verifies the operator circuit s high body voltage limit Figure 6 Pressing the...

Page 5: ...und 18 Set the variable DC power supply to 4 0 V The monitor s Iron LED should illuminate green and its audible alarm should stop 19 Disconnect the variable DC power supply from the monitor and ground Connect its positive terminal to ground and its negative terminal to the monitor s yellow alligator cord 20 Verify that the variable DC power supply is still set to 4 0 V The monitor s Iron LED shoul...

Page 6: ...e 12 Pressing the WRIST STRAP and HIGH BODY VOLTAGE test switches simultaneously 7 Disconnect the mono cable from the monitor VERIFYING THE MAT CIRCUIT 8 Connect the red test lead to red banana jack located at the top of the Verification Tester 9 Disconnect the monitor s white mat monitor cord from its worksurface mat and turn it over to expose its 10 mm snap 10 Clip the red test lead s mini grabb...

Page 7: ...ch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 20 Press and hold the Verification Tester s METAL GROUND PASS test switch Simultaneously press and hold the Verification Tester s EMI LOW test switch The monitor s tool LED will remain green and no audible alarm will sound This verifies the operator circuit s low EMI volt...

Page 8: ...s the tool circuit s impedance limit Figure 21 Pressing the METAL GROUND PASS test switch 5 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit Figure 22 Pressing the METAL GROUND FAIL test switch 6 Press and hold the Verification Tester s METAL GROUND PAS...

Page 9: ...necting the Verification Tester to the Ground Man Plus Workstation Monitor s operator jack Figure 26 Pressing the WRIST STRAP test switch Figure 27 Pressing the WRIST STRAP and LOW BODY VOLTAGE test switches simultaneously 4 Press and hold the Verification Tester s WRIST STRAP test switch The monitor s operator LED will illuminate green and its audible alarm will stop This verifies the operator ci...

Page 10: ...GROUND PASS test switch The monitor s tool LED will illuminate green and its audible alarm will stop This verifies the tool circuit s impedance limit 12 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 13 Press and hold the Verification Tester s METAL G...

Page 11: ... Battery LED illuminates red Open the compartment located on the back of the tester to replace the battery The tester uses one 9V alkaline battery Ensure that the battery s polarities are oriented correctly to avoid possible circuit damage Specifications Operating Temperature 50 to 95 F 10 to 35 C Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humid...

Reviews: