12. Press and hold the Verification Tester's SOFT 

GROUND test switch. The monitor's mat LED will 

illuminate green, and its audible alarm will stop 

after approximately 3 seconds. This verifies the mat 

circuit's resistance limit. 

Figure 8. Pressing the SOFT GROUND test switch

13. Disconnect the red test lead from the monitor's 

white mat monitor cord. 

14. Reinstall the white mat monitor cord to the 

worksurface mat.

VERIFYING THE IRON CIRCUIT

Note: A variable DC power supply must be used to 

complete this procedure. The CTE701 Verification 

Tester cannot verify the iron circuit in the Iron Man® 

Plus Workstation Monitor. 
15. Turn the voltage alarm trimpot at the back of the 

monitor fully clockwise. This configures it to ±5 V. 

16. Power the variable DC power supply. Configure it to 

5.0 V. 

17. Connect the negative terminal from the variable 

DC power supply to ground. Connect its positive 

terminal to the yellow alligator cord connected to the  

monitor's BOARD terminal. The monitor's Iron LED 

should illuminate red and its audible alarm should 

sound. 

18. Set the variable DC power supply to 4.0 V. The 

monitor's Iron LED should illuminate green and its 

audible alarm should stop. 

19. Disconnect the variable DC power supply from the 

monitor and ground. Connect its positive terminal 

to ground and its negative terminal to the monitor's 

yellow alligator cord. 

20. Verify that the variable DC power supply is still set 

to 4.0 V. The monitor's Iron LED should illuminate 

green. 

21. Set the variable DC power supply to 5.0 V. The 

monitor's Iron LED should illuminate red and its 

audible alarm should sound. 

WS Aware Monitor

CONFIGURING THE VERIFICATION TESTER

Configure the Verification Tester's DIP switch to the 

settings shown below. This will make its test limits 

match the factory default limits of the monitor.

VERIFYING THE OPERATOR CIRCUIT

1.  Use the black test lead to connect the Verification 

Tester to equipment ground. 

2.  Power the Verification Tester ON. 
3.  Use the 3.5 mm mono cable to connect the 

Verification Tester to the monitor's operator jack. 

The monitor's operator LED will illuminate red, and 

its alarm will sound. 

Figure 9. Connecting the Verification Tester to the WS 

Aware Monitor's operator jack

4.  Press and hold the Verification Tester's WRIST 

STRAP test switch. The monitor's operator LED will 

illuminate green, and its audible alarm will stop. This 

verifies the operator circuit's impedance limit. 

Figure 10. Pressing the WRIST STRAP test switch

Summary of Contents for CTE701

Page 1: ...nds annual calibration of workstation monitors and the CTE701 Verification Tester The CTE701 Verification Tester meets ANSI ESD S20 20 and Compliance Verification ESD TR53 November 2022 The SCS CTE701 Verification Tester can be used with the following items Item Description 770067 WS Aware Monitor 770068 WS Aware Monitor CTC061 3 242 WW WS Aware Monitor CTC061 RT 242 WW WS Aware Monitor CTC062 RT ...

Page 2: ...operator circuit when pressed F Soft Ground Test Switch Simulates a MAT PASS condition on the monitor when pressed G Wrist Strap Test Switch Simulates a OPERATOR PASS condition on the monitor when pressed H Test Limit DIP Switch Configures the test limits on the CTE701 Verification Tester I High Metal Ground Test Switch Simulates a TOOL FAIL condition on the monitor when pressed J High EMI Test Sw...

Page 3: ...FF 14 ohms OFF ON OFF OFF 15 ohms ON OFF OFF OFF 16 ohms OFF OFF OFF OFF EMI The EMI high frequency signal is configured with switch 9 The CTE701 Verification Tester provides two different levels of high frequency signal elevated and normal Pressing the HIGH EMI pushbutton will load a high signal level within its range Pressing the LOW EMI pushbutton will load a low signal within its range Switch ...

Page 4: ...OLTAGE test switches simultaneously 6 Continue to press and hold the Verification Tester s WRIST STRAP test switch Simultaneously press and hold the Verification Tester s HIGH BODY VOLTAGE test switch The monitor s green operator LED will illuminate continuously its red LED will blink and an audible alarm will sound This verifies the operator circuit s high body voltage limit Figure 6 Pressing the...

Page 5: ...und 18 Set the variable DC power supply to 4 0 V The monitor s Iron LED should illuminate green and its audible alarm should stop 19 Disconnect the variable DC power supply from the monitor and ground Connect its positive terminal to ground and its negative terminal to the monitor s yellow alligator cord 20 Verify that the variable DC power supply is still set to 4 0 V The monitor s Iron LED shoul...

Page 6: ...e 12 Pressing the WRIST STRAP and HIGH BODY VOLTAGE test switches simultaneously 7 Disconnect the mono cable from the monitor VERIFYING THE MAT CIRCUIT 8 Connect the red test lead to red banana jack located at the top of the Verification Tester 9 Disconnect the monitor s white mat monitor cord from its worksurface mat and turn it over to expose its 10 mm snap 10 Clip the red test lead s mini grabb...

Page 7: ...ch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 20 Press and hold the Verification Tester s METAL GROUND PASS test switch Simultaneously press and hold the Verification Tester s EMI LOW test switch The monitor s tool LED will remain green and no audible alarm will sound This verifies the operator circuit s low EMI volt...

Page 8: ...s the tool circuit s impedance limit Figure 21 Pressing the METAL GROUND PASS test switch 5 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit Figure 22 Pressing the METAL GROUND FAIL test switch 6 Press and hold the Verification Tester s METAL GROUND PAS...

Page 9: ...necting the Verification Tester to the Ground Man Plus Workstation Monitor s operator jack Figure 26 Pressing the WRIST STRAP test switch Figure 27 Pressing the WRIST STRAP and LOW BODY VOLTAGE test switches simultaneously 4 Press and hold the Verification Tester s WRIST STRAP test switch The monitor s operator LED will illuminate green and its audible alarm will stop This verifies the operator ci...

Page 10: ...GROUND PASS test switch The monitor s tool LED will illuminate green and its audible alarm will stop This verifies the tool circuit s impedance limit 12 Press and hold the Verification Tester s METAL GROUND FAIL test switch The monitor s tool LED will illuminate red and its audible alarm will sound This verifies the tool circuit s impedance limit 13 Press and hold the Verification Tester s METAL G...

Page 11: ... Battery LED illuminates red Open the compartment located on the back of the tester to replace the battery The tester uses one 9V alkaline battery Ensure that the battery s polarities are oriented correctly to avoid possible circuit damage Specifications Operating Temperature 50 to 95 F 10 to 35 C Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humid...

Reviews: