Troubleshooting the Device
51
SCOTT SAFETY
087-0038 REV L
Troubleshooting the
Device
Refer to
for troubleshooting assistance.
Table 13
Troubleshooting Matrix
CAUSE
SOLUTION
Low Flow Alarm
•
Check for obstructions, kinks in the hose, or other blockage
•
Verify no moisture has entered the hose or pump
•
Unscrew probe assembly and replace filter(s) in probe
•
Clean pump. See
“Cleaning the Pump” on page 46.
Unexpected chattering
noise coming from pump
during operation
•
Verify no moisture has entered the hose or pump
•
Clean pump. See
“Cleaning the Pump” on page 46.
•
Replace pump
Pump does not run when
activated
•
Verify no other alarms are active. Active alarms may inhibit
pump from powering on
•
Replace pump
The pump fails the
start-up flow test
•
Clean pump. See
“Cleaning the Pump” on page 46.
The sample draw reading
indicates less than
300cc/min (0.3LPM)
•
Clean pump. See
“Cleaning the Pump” on page 46.
Device within 1 to 2’ of a
high-powered portable
radio (reading fluctuates,
alarms & vibrates)
•
Perform Bump Test prior to putting device back into service.
Summary of Contents for Protege
Page 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 2: ......
Page 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Page 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Page 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Page 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Page 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Page 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Page 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Page 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Page 103: ......