TABLE
A-2: MODULE SELECTION TABLE ACCORDING TO STROKE VOLUME AND APPROXIMATE
SUBJECT WEIGHT
MODULE
STROKE
VOLUME
(mL)
MINIMUM
1
WEIGHT
(g)
MAXIMUM
1
WEIGHT (g)
FOR LARGE
AMPLITUDE
2
MANOEUVRES
MAXIMUM
1
WEIGHT (g) FOR
VENTILATION
AND SMALL
AMPLITUDE
3
MANOEUVRES
Y-TUBING
PORT SIZE
(in)
INTAKE AND
EXHAUST
PORT SIZE
(in)
4
FX1
1.53
8
30
150
1/8
1/8 (barbed)
FX2
4.22
20
85
420
3/16
1/8 (barbed)
FX3
12.4
60
240
1,230
3/16
1/8 (barbed)
FX4
28.2
150
560
2,810
3/8
1/4 (barbed)
FX5
51.7
260
1,000
5,100
3/8
1/4 (barbed)
Note:
1
Minimum and maximum weight ranges are subject to change with further development and
are listed for healthy subjects.
2
Estimated using 50 mL/kg ratio.
3
Estimated using 10 mL/kg ratio.
4
Intake and exhaust ports use Colder PMC/PLC connectors. FX Modules are shipped with
mating fittings for easy connection to standard tubing size
Summary of Contents for flexivent