68
Prior to the next use, check for cracks or damage, and replace the unit if any defects are
visible.
5.5.2
AERONEB CONTROLLER
1)
Wipe clean with a damp cloth.
2)
Check for exposed wiring, damaged connectors, or other defects and replace if any are
visible.
Do not autoclave the Aeroneb controller.
Do not immerse it in water.
5.6
ANNUAL CLEANING AND MAINTENANCE
The
flexiVent
modules have been designed to be field serviceable. With an inspiratory and
expiratory cartridge, the module was designed to minimize the downtime required for annual
cleaning and maintenance by allowing users to simply replace a cartridge in the comfort of their
own lab in a matter of minutes. We recommend that the expiratory cartridge be replaced once a
year or as needed based on usage. Please contact SCIREQ Customer Support to purchase spare
cartridges.
Note that we also offer a service plan that includes pre-paid cartridges along with an extended
warranty and software subscription. Please contact SCIREQ Customer Support for details.