59
Instruction Manual
– FB and FC Series Force Gauges
4.
Use the navigation keys to select Load-cell and press the Enter key
5.
Use the navigation keys to key in the MAX value of the load-cell and press the Enter key to continue.
6.
Use the navigation keys to key in the mV/V SENS (sensitivity) value of the loadcell and press the Enter key
.
MAX and mV/V values are located on the load-cell nameplate or Calibration Certificate.
CALIBRATION
1. Calibration START
2. Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3. Force: < 200 N, … >
4. g [ m/s2]: < 9.81416 >
5. Geogr. Location < Latitude: 54
°21’, Elevation 114m >
6. Correction < + 0.90N >
7. Load-Cell < MAX = 0 kg, SENS=0.00000 mV/V >
8. Factory Calibr. < Yes, No >
9. PIN < ****** >
10. Exit
CALIBRATION
1. Calibration START
2. Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3. Force: < 200 N, … >
4. g [ m/s2]: < 9.81416 >
5. Geogr. Location < Latitude: 54
°21’, Elevation 114m >
6. Correction < + 0.90N >
7. Load-Cell < MAX = 0 kg, SENS=0.00000 mV/V >
8. Factory Calibr. < Yes, No >
9. PIN < ****** >
10. Exit
Summary of Contents for TORBAL FB Precision Series
Page 1: ...Instruction Manual TORBAL FORCE GAUGES FB Precision FC Precision PRO Series...
Page 75: ...73 Instruction Manual FB and FC Series Force Gauges...
Page 76: ...74 Instruction Manual FB and FC Series Force Gauges...
Page 77: ...75 Instruction Manual FB and FC Series Force Gauges...
Page 78: ...76 Instruction Manual FB and FC Series Force Gauges...
Page 79: ...77 Instruction Manual FB and FC Series Force Gauges...
Page 80: ...78 Instruction Manual FB and FC Series Force Gauges...
Page 81: ...79 Instruction Manual FB and FC Series Force Gauges...
Page 82: ...80 Instruction Manual FB and FC Series Force Gauges...
Page 83: ...81 Instruction Manual FB and FC Series Force Gauges...