53
Instruction Manual
– FB and FC Series Force Gauges
13.4 Setting a Correction Value
1.
Position the device upside down with the hook tip attached. The gauge should be secured into a stand.
2.
Place the calibration load on the hook or hook-held platform.
3.
Note the difference between the known calibration load and the measured load.
4.
Select
Correction
and key in the noted difference.
STANDARD
MAN
AUT SDM
200.90 N
- +
STANDARD
MAN
AUT SDM
200.90 N
- +
CALIBRATION
1. Calibration START
2. Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3. Force: < 200 N, … >
4. g [ m/s2]: < 9.81416 >
5. Geogr. Location < Latitude: 54
°21’, Elevation 114m >
6. Correction < + 0.90N >
7. Load-Cell < MAX = 0 kg, SENS=0.00000 mV/V >
8. Factory Calibr. < Yes, No >
9. PIN < ****** >
10. Exit
CALIBRATION
1. Calibration START
2. Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3. Force: < 200 N, … >
4. g [ m/s2]: < 9.81416 >
5. Geogr. Location < Latitude: 54
°21’, Elevation 114m >
6. Correction < + 0.90N >
7. Load-Cell < MAX = 0 kg, SENS=0.00000 mV/V >
8. Factory Calibr. < Yes, No >
9. PIN < ****** >
10. Exit
Summary of Contents for TORBAL FB Precision Series
Page 1: ...Instruction Manual TORBAL FORCE GAUGES FB Precision FC Precision PRO Series...
Page 75: ...73 Instruction Manual FB and FC Series Force Gauges...
Page 76: ...74 Instruction Manual FB and FC Series Force Gauges...
Page 77: ...75 Instruction Manual FB and FC Series Force Gauges...
Page 78: ...76 Instruction Manual FB and FC Series Force Gauges...
Page 79: ...77 Instruction Manual FB and FC Series Force Gauges...
Page 80: ...78 Instruction Manual FB and FC Series Force Gauges...
Page 81: ...79 Instruction Manual FB and FC Series Force Gauges...
Page 82: ...80 Instruction Manual FB and FC Series Force Gauges...
Page 83: ...81 Instruction Manual FB and FC Series Force Gauges...