49
Instruction Manual
– FB and FC Series Force Gauges
To access the calibration menu, press the Menu key, select Calibration and press the Enter key.
13.1 Calibrating with Mass or Force
1.
Position the device upside down with the hook tip attached. The gauge should be secured into a stand.
2.
Select Mode and the unit of measure (LOAD or FORCE)
Chapter 13: Calibration
Chapter 14: Calibration
USER MENU
1. Applications
2. Data Manager
3. Measur. Settings
4. Configuration
5. Calibration
6. Info
7. Exit
USER MENU
1. Applications
2. Data Manager
3. Measur. Settings
4. Configuration
5. Calibration
6. Info
7. Exit
CALIBRATION
1. Calibration START
2.
Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3.
Force: < 200 N, … >
4. g [ m/s2]: < 9.81416 >
5. Geogr. Location < Latitude: 54
°21’, Elevation 114m >
6. Correction < + 0.00N >
7. Load-Cell < MAX = 0 kg, SENS=0.00000 mV/V >
8. Factory Calibr. < Yes, No >
9. PIN < ****** >
10. Exit
CALIBRATION
1. Calibration START
2.
Mode: < N/kN, kgf, lbf, ozf, kg, lb, oz >
3.
Force: < 200 N, … >
Summary of Contents for TORBAL FB Precision Series
Page 1: ...Instruction Manual TORBAL FORCE GAUGES FB Precision FC Precision PRO Series...
Page 75: ...73 Instruction Manual FB and FC Series Force Gauges...
Page 76: ...74 Instruction Manual FB and FC Series Force Gauges...
Page 77: ...75 Instruction Manual FB and FC Series Force Gauges...
Page 78: ...76 Instruction Manual FB and FC Series Force Gauges...
Page 79: ...77 Instruction Manual FB and FC Series Force Gauges...
Page 80: ...78 Instruction Manual FB and FC Series Force Gauges...
Page 81: ...79 Instruction Manual FB and FC Series Force Gauges...
Page 82: ...80 Instruction Manual FB and FC Series Force Gauges...
Page 83: ...81 Instruction Manual FB and FC Series Force Gauges...