21
Instruction Manual
– FB and FC Series Force Gauges
Standard Mode Configuration Features
Feature
Options
Description and Functionality
Status
<Enable, Disable>
Enables and Disables the standard mode from the function menu. When
disabled feature is not available for selection while pressing the function
key.
Mode
<Auto, Manual>
Manual:
Measurements are captured with the press of the start key.
Measurement sequence is based on the quantity parameter set in the
Quantity field.
Auto:
Measurements are automatically captured over specified period which
is based on the measurement time parameters. Pressing the start key
initiates the measurement cycle. The feature utilizes Delay and Trigger
settings.
Delay
<Start, Trigger>
Start:
Force trigger is disabled. The measurement sequence initiates after
press of the Start key. Measurement sequence begins once the Delay Time
value is reached.
Trigger:
Both Trigger and Delay Time values must be reached before
measurement sequence begins.
Trigger
custom value
The amount of Force that needs to be reached before the measurement
sequence begins. (Auto only)
Del. Time
custom value
The time that must be reached before the measurement sequence begins.
(Auto only)
Quantity
custom value
Number of measurements that must be captured to end the measurement
cycle. (Manual Only)
Meas. Time
custom value
Total time of the measurement sequence in the automatic mode.
Threshold
<On, Off>
Enables and disables measurement threshold monitoring.
Min
custom value
Minimum threshold value. When reached, MIN is displayed.
Target
custom value
Target threshold value when reached OK, is displayed.
Max
custom value
Maximum threshold value. When reached, and exceeded MAX is displayed
Exit
Exits the Applications settings folder.
3.
To begin measurement, align the force gauge shaft to the object measured. If measuring by hand grasp the
device firmly. For compression push the force gauge forward with the attached tip pressing into the object. For
tension wrap the hook tip around the object and pull the force gauge away from the object.
Summary of Contents for TORBAL FB Precision Series
Page 1: ...Instruction Manual TORBAL FORCE GAUGES FB Precision FC Precision PRO Series...
Page 75: ...73 Instruction Manual FB and FC Series Force Gauges...
Page 76: ...74 Instruction Manual FB and FC Series Force Gauges...
Page 77: ...75 Instruction Manual FB and FC Series Force Gauges...
Page 78: ...76 Instruction Manual FB and FC Series Force Gauges...
Page 79: ...77 Instruction Manual FB and FC Series Force Gauges...
Page 80: ...78 Instruction Manual FB and FC Series Force Gauges...
Page 81: ...79 Instruction Manual FB and FC Series Force Gauges...
Page 82: ...80 Instruction Manual FB and FC Series Force Gauges...
Page 83: ...81 Instruction Manual FB and FC Series Force Gauges...