41
Command Description
Answer
99AAnn
Transmit new address (nn = new address)
nnY
aaAB
command
Command after instruction AB ... is performed
aaY
and sent to following device
aaBF
Fill piston burette
aaY
aaBN
Reset ml display to zero
aaY
aaBP
Inquiry of piston position in scale parts
aaBP=0 ... 8000
aaBV
Inquiry of burette volume
aaV=12.34
aaDA13.56
Dose with addition of volume
aaY
aaDO123.4
Dose indicated quantity in ml
aaY
aaDB123.4
Dose quantity in ml without filling and zero
aaY
aaDH
Endless dosing, to be stopped
aaDZ
Start dosing, without filling; display will add up to max. volume 9999 ml
aaY
aaGDM20,3
Dosing rate in ml/min
aaY
aaGF30
Filling time in s
aaY
aaKA
Which key is being pressed?
AaKey:
“Text“
aaKD
Keyboard disable
aaY
aaKE
Keyboard enable
aaY
aaMO0
Inquiry of current function
aaModeZ
Digit sequence: Z = 1
ð
„Titration“
Z = 2
ð
„Dosing“
Z = 3
ð
„Incremental dosing“
aaMO1
Set titration function
aaY
aaMO2
Set dosing function
aaY
aaMO3
Set incremental dosing function
aaY
aaRA
Interchangeable unit - inquiry of volume size
aaAUFSATZ:20
aaRH
Identification request
aaIdent:T universal
aaRC
Send last command
aa“last command“
aaRS
Report status
aaStatus:“text“
aaSC
Continue after stop/continue function after stop
aaY
aaSH
Stop, burette stops
aaY
aaSD1..8
Write string in n
th
matrix line: attach string to be written with “TEXT“
aaY
(limit characters to 21)
as a delimiter character to command e.g.: 01SD11”
(Please replace interchangeable unit)
aaST0
Delete display completely
aaY
aaST1
Switch display fully on
aaY
aaSR
Stop current function
aaY
aaSYS5
Set display language to German
aaY
aaSYS1
Set display language to English -english-
aaY
aaSYS2
Set display language to French -français-
aaY
aaSYS3
Set display language to Spanish -español-
aaY
aaVE
Software version number
aaVersion:99.23
aaVO
Volume inquiry from display
aaV=12.345ml