34
You can use the <
↑
>, <
→
>, <
↓
> arrow keys on the front panel of the TITRONIC
â
â
â
â
universal
Piston Burette
to set a filling time off min. 30 s to max. 999 s. The display will show your modified values. The cursor is lo-
cated below the digit to be changed. As soon as the cursor has reached the last digit to the right, it will
jump back to the first digit to the left if the <
→
> arrow key is pressed again, display: <<Next digit>>. The
filling time refers to the duration of one complete piston stroke during the drawing in off dosing or titration
solution. This time is increased by the switching times of the valves of approximately 6 s. If you trigger the
filling process with their contents of the cylinder not being fully ejected, "Filling" will be shortened corre-
spondingly.
The <F3> key serves to select the national language or to set the parameters of the RS-232-C interface.
After a waiting time of 15 seconds the burette will automatically switch to the last selected working method
(function), e.g. „Dosing“ = <Mode>:Dos. The selected working method remains stored after the burette is
switched off.
Display
Key
What is happening, and what is to be done?
Rinsing
<F1> Double filling and ejection of the cylinder contents
Filling time
<F2> Setting the filling time
System
<F3> Selection of the national language, and setting of the serial parameters
Menu
<Mode> Selection of the functions: Dosing, titration, and incremental dosing
4.3 Selection of the national language
For the display of the TITRONIC
â
â
â
â
universal
Piston Burette you can select one of the following national
languages: German, English, French, or Spanish. From the switch-on image,
&
Fig. 10, use the <F3> key
to switch to the language selection. Then you press the <F1> key for ‘Deutsch’, the <F2> key for ‘English’,
the <F3> key for ‘Français’, or the <F4> key for ‘Español’.
Deutsch <F1>
English <F2>
Français <F3>
Español <F4>
Pressing the <MODE / F4> key for approx. 3 s will get you to the switch-on image.
4.4 Rinsing and initial filling
Considering that a max. of 100 ml of dosing or titration solution will be ejected during “Rinsing/Initial filling”,
a vessel with a min. capacity of 100 ml has to be placed under the burette tip.
!
Please note
! :
Please wear protective glasses and a laboratory coat and observe the UVV1 and
UVV 113 (VBG) accident-prevention regulations as well as the pertinent
implementation comments.
Starting the function
Fig. 12
TITRONIC universal
Ver.: 99.14
Cylinder volume 20 ml
Rins./init. fill.
<F1>
Filling time 30 (s)
<F2>
System
<F3>
Menu
<Mode>
Press the <FILL> / <F1> key.
Fig. 13
To cancel the function: Press the <STOP> key.