Trouble shooting

The device does not respond to keys being pressed, the display is dark.

The internal program is disturbed
(e.g. by electrostatic charging or
mains overvoltage).

Switch device off, switch it on again after 10 s.

The display contrast is misadjusted.

Use the knurled knob at the right side
of the bottom to adjust the contrast.

The stirrer does not work.

The connection contacts are dirty.

Clean contacts.

The dosing unit is not filled correctly.

The reagent bottle is empty.

Replace or top up reagent bottle.

The hose is not immersed deeply
enough in the reagent bottle.

Immerse the hose deeper in the reagent bottle
or top up reagents.

The dosing unit is not properly locked.

Latch dosing unit with a quarter rotation in the lower position.

Air bubbles in the system

The valve is defective.

Replace valve.

The hose connections are not tight.

Check: Has the hose been pulled out of the
threaded connection? Fasten fingertight.
Replace hoses plus screw connectors.

The hose is not immersed deeply
enough in the reagent bottle.

Immerse the hose deeper in the reagent bottle
or top up reagents.

The titration tip drips after use.

Air bubbles in the system

.

Set longer filling time.

When installing the dosing unit the piston rod is not pulled in correctly, the device “rattles”.

Oblique application of piston rod.

Re-apply the piston rod vertically.

The titration solution is not titrated/dosed.

The dosing unit is not properly filled

Perform initial filling.

The dosing unit is not properly locked.

Lock dosing unit with a quarter rotation
in the lower position (

&

 Fig. 30).

The hose or the titration tip
are bent or obstructed.

Check proper flow through hose and tip,
replace if required.

The system contains air bubbles.

Please refer to “Air bubbles in the dosing system”.

Undissolved parts in the titration solution.

Filter or replace titration solution.

Valve hoses are mixed up.

Connect hoses (

&

 Fig. 1 + 8).

Data transfer to the printer/computer does not work.

Set proper transfer parameters.

Setting: 4800 baud, 7 data bits, no parity, no handshake.

Using an appropriate cable?

Use TZ 3095 for printer or TZ 3096 / TZ 3097 for PC.

Cable correctly connected?

The 

middle

 socket is the RS-232-C interface 1.

Check cable connection; are the screws on the side
of the PC tightened?
PC: Possibly „COM 1“ and „COM 2“ were mixed up!

Defective cable.

Replace cable.

Malfunction of printer/PC or

TITRONIC

â

â

â

â

 

universal  Piston Burette

Switch device off, switch in on again after 10 s.

Summary of Contents for TITRONIC universal

Page 1: ...sanleitung Kolbenb rette TITRONIC universal Operating Instructions Piston Burette TITRONIC universal Mode d emploi Burette piston TITRONIC universal Manual de instrucciones Bureta de mbolo TITRONIC un...

Page 2: ...iversal without changing the described properties Mode d emploi Page 49 72 Instructions importantes Pri re de lire et d observer attentivement le mode d emploi avant la premi re mise en marche de la B...

Page 3: ...pplier s Certificate We certify that the equipment EN ISO 9001 part 4 10 4 Final inspection and testing and that the specified requirements for the product are met Certificat du fournisseur Nous certi...

Page 4: ...TRONIC universal Piston Burette 33 4 1 Keys and contrast 33 4 2 Switch on image setting the filling time 33 4 3 Selecting the national language 34 4 4 Rinsing and initial filling 34 4 5 Filling 35 4 6...

Page 5: ...26 Fig 1 TITRONIC universal Piston Burette with TM 96 Stirrer with manual key button with stand with electrode burette tip holder with burette tip TZ 1503 with bottle top TZ 2005...

Page 6: ...contact with the titration and dosing solutions are made of chemically resistant mate rials Analysis accuracy is guaranteed by the high precision glass cylinder and the spindle for the piston feed Th...

Page 7: ...ieht bereinstimmt mit den Normen und mit dem normativen Dokument We declare under our sole responsibility that the product Piston Burette TITRONIC universal to which this declaration relates is in con...

Page 8: ...ading and 2 trailing decimal digits with a resolution of 0 01 ml Display Matrix LCD display 69 x 39 mm 64 x 128 pixels with background lighting contrast can be set using a knurled wheel Connections Ma...

Page 9: ...ONIC universal Piston Burette from the place of work Examples for the assumption that a safe operation is no longer possible the package is damaged the TITRONIC universal Piston Burette shows visible...

Page 10: ...he operating voltage of the TITRONIC universal Piston Burette matches the mains voltage The operating voltage is indicated on the specification plate on the bottom of the TITRONIC universal Piston Bur...

Page 11: ...off Mains cable Interface 2 RS 232 C Connector for manual key button Interface 1 RS 232 C The transfer parameters for the RS 232 C interface can be selected according to RS 232 C interface chapter Att...

Page 12: ...ered using the keyboard In dialogue mode the display shows which active keys can be pressed for further work The manual key button is used to start the START STOP and the FILL function and to output t...

Page 13: ...ction of the cylinder contents Filling time F2 Setting the filling time System F3 Selection of the national language and setting of the serial parameters Menu Mode Selection of the functions Dosing ti...

Page 14: ...al Piston Burette the display will show the respective basic menu of a function Here you use the FILL or the grey key of the manual key button to start the Filling function Upon completion of the manu...

Page 15: ...you can reset the input to the last valid parameters by pressing the F4 key Fig 16 Fig 17 Dosing parameter Nominal vol 015 00ml Change digit Next digit OK F1 ESC F4 Dosing parameter Speed 40 00ml min...

Page 16: ...wn including 2 nd stage high speed and keep it depressed The titration solution will be dosed as long as the key is kept depressed The display will show the volume and the selected rate stage with the...

Page 17: ...ith the display of the TITRONIC universal Piston Burette Setting the parameters chapter The stored parameters remain active even after the burette is switched off 4 9 Incremental dosing Use the MODE k...

Page 18: ...g with each other require the same transfer parameters The following transfer parameters are set ex factory Baud rate 4800 baud Data bits 7 Bit Stop bits 1 Parity No If you wish to change the data tra...

Page 19: ...ton Burette This piston burette in turn will now check the address and will respond to this command just like the first burette Any information data strings arriving at the interface 2 of the TITRONIC...

Page 20: ...ce Z 1 Titration Z 2 Dosing Z 3 Incremental dosing aaMO1 Set titration function aaY aaMO2 Set dosing function aaY aaMO3 Set incremental dosing function aaY aaRA Interchangeable unit inquiry of volume...

Page 21: ...2 or Fig 9 item 2 Important note The disassembly of the dosing unit as such is only possible using a special tool TZ 3630 and should only be done by experienced persons The sealing lip of the PTFE pis...

Page 22: ...the cylinder and in the hoses The disturbances to be anticipated are the greater the more the new solu tion differs from the previous type and concentration In the case of highly different solutions t...

Page 23: ...rips after use Air bubbles in the system Set longer filling time When installing the dosing unit the piston rod is not pulled in correctly the device rattles Oblique application of piston rod Re apply...

Page 24: ...30 If moisture is actually present one can conclude that the piston in the cylinder is no longer tight Check the hoses screwed connections and seals for visible damage contamination and leakage assem...

Page 25: ...r repair please state the titration solution last used on the accompanying papers or on the TITRONIC universal Piston Burette itself Otherwise the service department will dispose of the damaged or rep...

Page 26: ...uter PC with 25 channel plug 1 side 4 channel mini DIN plug 2 side 25 channel sub Miniature D plug TZ 3097 1 5 m data cable RS 232 C for computer PC with 9 channel plug 1 side 4 channel mini DIN plug...

Page 27: ...48 Order numbers Fig 31 M3 X 10 mm TZ 3660 TZ 1503 TZ 3460 TZ 3025 TZ 2008 TZ 2005 TZ 3095 TITRONIC universal TZ 3665 TZ 3680...

Reviews: