
Operation
ProLab 2000
92
ba75577e03
11/2009
4.10.4 Operation with MultiLab pilot
With the aid of the MultiLab pilot software, you can record and evaluate
measuring data with a PC. The data is transmitted after the meter was
connected to the RS232 serial interface (COM port) or the USB
interface of the PC.
Note
More detailed information can be found in the MultiLab pilot operating
manual.
4.11
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.11.1 Resetting the sensor settings
Note
The calibration data is reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH
The following settings for pH measurements are reset to the default
settings with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the
respective measuring menu. To open it, activate the relevant
measuring window in the measured value display and press the
<MENU/OK>
key.
Setting
Default settings
Cal. interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
High resolution pH
On
High resolution mV
On
Zero point
0 mV
Slope
-59.16 mV
Temperature, manual
25 °C