Overview
ProLab 2000
14
ba75577e03
11/2009
ID sensors
SI Analytics GmbH ID sensors support the automatic sensor
recognition function. Their sensor designation has the addition "ID",
e.g. A 161 1M DIN-ID electrode.
Note
Information on available ID sensors is given on the Internet or directly
by SI Analytics.
ID sensors connected to the ProLab 2000 meter can be recognized by
the ID sensor symbol on the display of the meter.
Sensor data from ID
sensors
ID sensors transmit the following sensor data to the meter:
z
SENSOR ID
– Sensor type
– Sensor series number
z
Calibration data
– Calibration date
– Calibration characteristics
– Calibration interval
– Selected buffer set (pH electrodes only)
The calibration data is updated in the ID sensor after each calibration
procedure. The ID sensor symbol flashes while this is done.
Note
While the ID sensor symbol is flashing, the sensor must not be
disconnected because otherwise the calibration data will not be
completely transmitted. The sensor will then have no valid calibration.
Note
If non-ID sensors are used, the calibration data from the meter is used
and also stored in the meter.
pH
7.000
25.0
°C
26.04.06 10:37
ID sensor symbol