ProLab 2000
Operation
79
ba75577e03
11/2009
5
Press
<MENU/OK>
.
The measurement with stability control to determine the D.O.
concentration starts.
6
Wait for the end of the measurement with Stability control or
accept the measured value with
<MENU/OK>
.
The determined D.O. content is displayed as a message (
i
).
7
Open the setting of the correction factor with
<MENU/OK>
.
8
Using
<
▲
><
▼
>, set the correction factor to adjust the
displayed concentration value to the nominal value (value of
the comparison measurement). Subsequently, accept the
correction factor with
<MENU/OK>
.
9
Select
Accept
with
<
▲
><
▼
> and press
<MENU/OK>
.
The meter switches to the measured value display. The
[Factor]
indicator appears in the measuring window.
i
Concentration =
8.5 mg/l
i
Temperature =
24.8 °C
Terminate stability control
LFOx1400-ID <CAL>
i
Concentration =
8.49 mg/l
i
Temperature =
24.8 °C
Set factor
:
1.000
Accept
LFOx1400-ID <CAL>