ProLab 2000
Operation
69
ba75577e03
11/2009
4
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
5
Press
<MENU/OK>
.
The measurement with stability control to determine the cell
constant starts.
6
Wait for the end of the measurement with Stability control or
accept the calibration value with
<MENU/OK>
.
The determined cell constant is displayed as a message (
i
).
i
Cell constant =
0.475 1/cm
i
Immerse sensor in solution
Continue
<CAL>
i
Cell constant =
0.481 1/cm
i
Temperature =
24.8 °C
Terminate stability control
<CAL>
i
Cell constant =
0.481 1/cm
i
++
Accept
<CAL>