
ProLab 2000
Operation
77
ba75577e03
11/2009
1
Put the D.O. sensor into the air calibration vessel.
2
Connect the D.O. sensor to the meter.
3
In the measured value display, select the D.O. measuring
window with
<
▲
><
▼
> and
<MODE>
.
4
Start the calibration with
<CAL>
.
The calibration display for the relevant sensor type appears.
5
Put the D.O. sensor into the air calibration vessel.
6
Press
<MENU/OK>
.
The measurement with stability control to determine the
relative slope starts.
7
Wait for the end of the measurement with Stability control or
accept the calibration value with
<MENU/OK>
.
The determined relative slope is displayed as a message (
i
).
i
Relative slope =
1.00
i
Use air calibration vessel
Continue
LFOx1400-ID <CAL>
i
Relative slope =
1.00
i
Temperature =
24.8 °C
Terminate stability control
LFOx1400-ID <CAL>