ProLab 2000
Operation
41
ba75577e03
11/2009
4.5.4
Settings for pH and ORP measurements
Overview
The following settings are possible for pH and ORP measurements:
z
Resolution
z
Calibration interval
z
Buffers for calibration
z
Unit for zero point
z
Unit for slope
z
Calibration record
(display)
Settings
The settings are made in the measuring menu of the pH/ORP
measurement. To open it, activate the relevant measuring window in
the measured value display and press
<MENU/OK>
. After completing
the settings, switch to the measured value display with
<MODE>
.
Menu item
Possible setting
Description
Calibration
/
Calibration record
-
Displays the calibration
record of the last
calibration.
Calibration
/
Buffer
NIST/DIN
TEC
MERCK 1
MERCK 2
MERCK 3
DIN 19267
Mettler US
Mettler EU
Fisher 1
Fluka BS
Radiometer
Baker
Metrohm
Beckmann
Hamilton DC
Precisa
Buffer sets to be used for
pH calibration.
For details, see page 43.
Calibration
/
Calibration interval
1 ... 999 d
Calibration interval
for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing CalClock in the
measuring window.
Calibration
/
Unit for
zero point
mV
pH
Unit for the zero point.