AR100 Operation Manual
© 2018 Schmitt Industries
Revision 1.1
26
maximum sampling frequency of the sensor (9.4 kHz) and this value increases in
proportion to the frequency reduction.
Voltage output 0…10 V
The connection scheme is shown in the figure. To reduce noise, it is
recommended to install RC filter before the measuring instrument. The filter
capacitor value is indicated for maximum sampling frequency of the sensor (9.4
kHz) and this value increases in proportion to the frequency reduction.
Configuration parameters
Range of the analog output
While working with the analog output, resolution can be increased by using the
‘Window in the operating range’ function which makes it possible to select a
window of required size and position in the operating range of the sensor within
which the whole range of analog output signal will be scaled.