AR100 Operation Manual
© 2018 Schmitt Industries
Revision 1.1
17
1.1.1. List of parameters
Code of
paramet
er
Description
Values
00h
Sensor ON
1 — laser is ON, measurements are taken
(default);
0 — laser is OFF, sensor in power save
mode.
01h
Analog output ON
1 — analog output is ON;
0 — analog output is OFF.
02h
Control of averaging,
sampling, AL and analog
output
x,M2,A,С,M1,M0,R,S – control byte which
determines the operation mode.
M2:M1:M0
bits (AL mode):
000 - out of the range indication (by
default);
001 - 'slave' mode (mutual
synchronization);
010 - hardware zero set mode;
011 - laser switch OFF/ON;
100 - encoder mode;
101 - input mode;
110 - reset of the Ethernet packets
counter;
111 - 'master' mode (mutual
synchronization).
A
bit (averaging mode):
0 - averaging over a number of
results (by default);
1 - time averaging (5 ms).
C
bit is not used
R
bit (analog output mode):
0 - window mode (by default);
1 - full range.
S
bit (sampling mode):
0 - time sampling (by default);
1 - trigger sampling.
x
bit is not used
03h
Network address
1…127 (default — 1)
04h
Rate of data transfer
through a serial port
1…192 (default — 4), specifies data
transfer rate in increments of 2400 baud;
e.g., 4 means the rate of 4*2400=9600
baud.
05h
Reserved
06h
Number of averaged values 1…128 (default — 1)