Safety Notes 

 

       

  

General References 

 
The enclosed operating instruction is intended for technically qualified personnel who has corresponding knowledge in the 
field of measurement and industrial process & control technology. 
The precise information about all safety notes contained in this operation manual and warnings, as well as its perfect 
technical implementation are precondition for the safe installation, the initiation, the secure operation and the maintenance 
of technology devices. For this purpose it is absolutely necessary that all measures are carried out by qualified personnel. 
All persons concerned with the project planning, installation and service of devices, must be familiar to the security concepts 
in automatic control and should be qualified in this sense. 
For clarity reasons, the enclosed operating instruction can not represent complete details in all conceivable cases of 
applications for the handling of devices. Further, we cannot consider the entire types of installation, handling and 
maintenance. If you wish further information or if special problems occur, which were not ,or not at length represented in this 
operating instruction, contact us, please.  
 
The oblivion of the safety notes can and will lead to material damages, body injuries and death. 
 

The devices may only be operated in accordance with the applications described in this operation manual. Built-in 
devices may only be operated in appropriate installations.   

 
With the connection and the initiation of the device, the customer accepts the general sale and delivery conditions. Further, 
he accepts eventually incomplete operation manuals. The information described is without guarantee. Errors and changes 
are reserved.             
   
Intended Purpose, Improper Usage 
   
A device is used for displaying, processing and controlling or regulation of processes. It shall not be used as the only tool for 
the prevention of dangerous states to machines and plants. Machines and plants must be constructed in such a way, that 
erroneous states can not lead to a dangerous situation for the staff (e.g. by independent limit switches, mechanical locking 
devices). It must be guaranteed in particular that device-operating errors, its malfunction or it’s breakdown do not lead to 
great property damages or danger for the staff. Consequently, the device then can be used to prevent the machine or the 
technical installations from error conditions.  
It is also important that the use of devices does not endanger precautions for the safety of technical installations. 
Emergency-off settings must remain effective in all operation modes. 
  
Installation Notes  
  
The

 

devices

 

must be installed and connected by compliance with the relevant DIN- and VDE-norms. They must be installed 

in such manner that an unintentional use is sufficingly excluded. The corresponding hardware and software safety 
precautions are to be observed in such manner that an interruption of the supply and signal cables cannot lead to an 
undefined or dangerous state. Supply and signal cables must be installed in such way, that disturbing signals (e.g. inductive 
or capacitive intersperses) will not cause derogations to the function of devices. 
 
Disturbance, Maintenance and Repair Notes 
   
The devices do not contain parts which can be maintained on the customer side. Repairs shall be carried out by exclusively. 
If assuming that a safe operation of the device is not possible anymore, it must be closed down and protected against 
unintentional handling immediately. This, in particular, applies: 
 

If the device shows visible damages 
If the device is no longer operative    
If parts of the device are loose or slack  
If the connection cables show visible damages   

 

Furthermore, we point out that all obligations of exclusively result from the respective sales contract in which the 
guarantee has been conclusively settled.

Summary of Contents for GM 80 PA

Page 1: ...Operating instruction DC Voltage Measuring Amplifier with Data Logger GM 80 TG PA V1 __ ...

Page 2: ...f processes It shall not be used as the only tool for the prevention of dangerous states to machines and plants Machines and plants must be constructed in such a way that erroneous states can not lead to a dangerous situation for the staff e g by independent limit switches mechanical locking devices It must be guaranteed in particular that device operating errors its malfunction or it s breakdown ...

Page 3: ...parameter set is documented on the rear of the device Operation Example GM 80 PA after current supply is immediately in the measuring mode of the previous selected parameter set GM 80 TG is switched on by pressing ENTER Key Assignment Key Menu Scroll up Scroll down ENTER Confirmation one step up in the menu ESC Discard one step back in the menu Control time and date in SYSTEM menu or adjust if nec...

Page 4: ...sured value shall be sent through an interface to a pc or printer the baud rate must be adjusted to the receiver and at SCI MODE the operating mode must be chosen For example HAND a measured value with time is displayed during every keystroke on ENTER while measuring For data logging adjust operating mode in the logger mode e g choose AUTO and interval time e g 10 sec and change to measuring mode ...

Page 5: ...GM 80 TG PA Brief Handling Overview ...

Page 6: ... It is an optocoupler input and therefore it is separated galvanically As a control signal 0V 2V is recognized as low level 3 5V 27V as high level Trigger Thus depending on the adjustment data can be stored or interface data can be issued This input has a high sampling rate therefore very short impulses are recognized as well Enable By this the min and max value is deleted externally In the displa...

Page 7: ...0 s at very slow courses a small rate e g 1 s AVERAGE Forming an average value the numbers indicate by how many measurements an average value is accumulated Applicable at e g vibrations control oscillations 2 5 PASS Password query After entering the password 9373 it is possible to enter menu 3 CALIB There the sensor parameters can be modified 2 6 LCD LCD contrast adjustment If the LCD becomes unre...

Page 8: ...is active GWS 1 2 GWS 1 and 2 are active GWS2 is activated via control input 3 CALIB Calibration menu this is active only if the password has been entered correctly in 2 5PASS For the calibration the sensor needs to be connected to the sensor socket Sensor__ Sensor number or name of the sensor which is to be calibrated POINTDIS Measuring range with decimal indication of the sensor e g 100 0 200 0 ...

Page 9: ...0 LOAD unload sensor 100 LOAD calibration by 100 load apply the nominal load nominal torque or switch to nominal value or NOMVALUE fixed value 4 20 mA SAVE query for takeover of calibration data Edit and Adjust Limit Values In menu 2 SYSTEM 2 9 INPUT Setpoint the limit value parameters can be adjusted If menu LIMIT VALUE SETS is selected and confirmed by ENTER the limit values are listed Switch to...

Page 10: ...play 4 0x34 Reset maximal value 5 0x35 Reset minimal value 6 0x36 Actuate calibration control for sensors with 100 control resistance 7 0x37 Switch off calibration control for sensors with 100 control resistance A 0x41 Read Out datalogger B 0x42 Delete datalogger C 0x43 Read Out current sensor parameter D 0x44 Read Out status E 0x45 Read Out complete status g 0x67 Change protocol setup k 0x6B ENTE...

Page 11: ...dit 0 load mA and 100 load mA Dec point 0xXXXX XXXX 000 _5000___ 001 _5 000__ 010 _50 00__ 011 _500 0__ 100 5 000___ 0 load 2Byte HEX value MSB LSB 100 load 2Byte HEX value MSB LSB Read out of status Status 2Byte general error condition of the GM80 Read out of complete status Status 2Byte general error condition of the GM80 Meas rate 1Byte 0x01 1000 sec 0x02 100 sec 0x03 10 sec 0x04 1 sec Average ...

Page 12: ...input is ended either if 256 characters are received or if the character ETX 0x03 Strg C is contained in the character string Change of Protocol Setup Protocol status 1Byte 0xXXXX XXXX binary coded 1 no closing character is sent 1 send CR LF 1 send CR 1 send LF Calculation of the Checksum CS and the Weighted Checksum gewCS The calculation takes place via all parameter bytes without the command byt...

Page 13: ...In this mode a measured value is issued via the serial interface when the Enter key is pressed at the GM 80 AUTO In this mode a measured value is issued via the serial interface in the adjusted delay TRIGGER At a trigger event In this mode a measured value is written in the logger Since the trigger pulses may occur in 10ms raster and additional time logg is not possible The flank of the trigger pu...

Page 14: ...starting time Since in GRAPHS and in the WINDOW MODE trigger events smaller than 1sec can occur an additional time log is not possible Adjustments in the Logg Mode LOGG OUT Here the data logger is switched off The logg mode in the measuring mode is switched to LOGG OUT as soon as the entire data logger has been edited HAND In this mode a measured value is written into the data logger when the ente...

Page 15: ...Pin Assignment ...

Page 16: ...attery accumulator inserted correctly Connect power supply at operation with accumulators The display begins to flash Charge the accumulators and or exchange the batteries The calibrated adjustments do not match with the sensor anymore Was the sensor overloaded Is the correct sensor selected Possibly recalibrate the sensor Wrong values are indicated after the calibration Was the sensor type select...