Menu Description:
1 MEASURE:
Measuring mode
SENSOR__:
Sensor selection for measuring mode, sensor 0 - 9
Here the sensor parameter set for the measurement is being selected.
The sensor parameter set must match the connected sensor.
2 SYSTEM:
All system parameters are being stored in this menu column.
2.1 LANG
Language adjustment
The menu language can be chosen in German, English, French or Spanish.
2.2 INFO
Information query was configured factory-sided and is not variable.
VERS NO.
Software version
2.3 SCI
Interface configuration
BAUD:
Baud rate adjustment must be conform with the receiver (pc or printer).
SCI MODE
Interface configuration only refers to the measuring mode, not the logg mode!
OFF
Interface
off
HAND
At key stroke "ENTER“ the measured value is issued
AUTO
Interval time, adjustable from 10ms to 1h (at high measuring rate only possible with highest baud
rate)
TRIGGER
At a trigger pulse a measured value is issued
TRIGGER2
At a trigger pulse a measured value is issued and the MIN-MAX-value is deleted.
deleted.
ADDRESS
Address
selection
2.4 RATE
Measuring rate and average value
RATE
Measuring rate adjustment can be chosen between 1/s and 1000/s. At fast courses (screw joint,
insert press...) always choose a fast measuring rate, e.g. 1000/s, at very slow courses a small rate,
e.g. 1/s.
AVERAGE
Forming an average value, the numbers indicate by how many measurements an average value is
accumulated. Applicable at e.g. vibrations, control oscillations.....
2.5 PASS
Password query
After entering the password
9373
it is possible to enter menu
3 CALIB
.
There the sensor parameters can be modified.
2.6 LCD
LCD-contrast adjustment
If the LCD becomes unreadable by external influences e.g. solar radiation, heat or cold, the LCD-
contrast can be corrected here.
2.7 LOGG
Data logger adjustments
LOGGMODE
Data
logger
configuration.
OFF
Off-switch for logger operation, by this the measuring mode is possible only.
HAND
At each key stroke on Enter a logg value is being taken over.
AUTO
Adjustment of the measuring intervals.
DIAGRAM With each increasing flank of the trigger signal a logg value is being taken over.
WINDOW The window operation is being started by an increasing trigger flank and can only be finished by a
decreasing flank. During this time, measured values will be deposited in the data logger in 1ms
raster.
TRIGGER2
With each increasing flank of the trigger signal a logg-value is taken over and the MIN-
MAX-value is deleted.
LOGGSEND
Logger values are issued through the interface, press ENTER twice
LOGGDEL
Logger values are deleted, beforehand there will be a safety query whether the buffer should be
deleted. Before deleting please assure that you have received the desired data.