Single slope adjustment
Allows acquiring stable measurement with a load on the scale to set the first segment. Performing physical
calibration in this way do not allows multiple segments calibration. Use this calibration method only if you
cannot acquire the segment calibration immediately after the Zero calibration (e.g.: big hopper). Zero
calibration must have been done before.
The steps of this single slope adjustment operation are the followings :
1.
Define the number of points to realize the calibration and the load values for each,
2.
Press
,
3.
Place the Load 1 and
press
,
4.
Wait for the end of data acquisition,
5.
Repeat steps 3 and 4 (if several points have been defined).
Remark
: If you press
,
the operation will stop and the values will not be saved.
The latest valid calibration values will be restored.
Theoretical calibration
An automatic scaling to migrate from the factory calibration to the user calibration.