Instruction Manual 

HMP 

 

HMP-BA-e-1211 

 

9

 

Super strength steel 

630

800 

1180

2652 

10 

Stainless steel 

500

710 

703

1676 

 

    
 
 
 

 
Table 3 
 

Type of impact device 

DC(D)/DL 

D+15 

Impact energy 
Mass of impact body 

11mJ 
5.5g/7.2g 

11mJ 
7.8g 

2.7mJ 
3.0g 

90mJ 
20.0g 

11mJ 
5.5g 

Test tip hardness:  
Diam. Test tip: 
Material of test tip: 

1600HV 
3mm 
Tungsten 
carbide 

1600HV 
3mm 
Tungsten 
carbide 

1600HV 
3mm 
Tungsten carbide 

1600HV 
5mm 
Tungsten carbide 

5000HV 
3mm 
synthetic diamond 

Impact device diameter: 
Impact device length: 
Impact device weight: 

20mm 
86(147)/  75mm 
50g 

20mm 
162mm 
80g 

20mm 
141mm 
75g 

30mm 
254mm 
250g 

20mm 
155mm 
80g 

Max. hardness of sample 

940HV 

940HV 

1000HV 

650HB 

1200HV 

Mean roughness value of  
sample surface Ra: 

1.6μm 

1.6μm 

0.4

μm 

6.3μm 

1.6μm 

Min.weight of sample: 
Measure directly 
Need support firmly 
Need coupling tightly 

 
>5kg 
2

5kg 

0.05

2kg 

 
>5kg 
2

5kg 

0.05

2kg 

 
>1.5kg 
0.5

1.5kg 

0.02

0.5kg 

 
>15kg 
5

15kg 

0.5

5kg 

 
>5kg 
2

5kg 

0.05

2kg 

Min. Thickness of sample : 
Coupling tightly 
Min. layer thickness for  
surface hardening 

 
5mm 
 
≥0.8mm 

 
5mm 
 
≥0.8mm 

 
1mm 
 
≥0.2mm 

 
10mm 
 
≥1.2mm 

 
5mm 
 
≥0.8mm 

Size of tip indentation 

Hardness 
300HV 

Indentation 
diameter 
Depth of indentation 

0.54mm 
 
24

μm 

0.54mm 
 
24

μm 

0.38mm 
 
12

μm 

1.03mm 
 
53

μm 

0.54mm 
 
24

μm 

Hardness 
600HV 

Indentation 
diameter 
Depth of indentation 

0.54mm 
 
17

μm 

0.54mm 
 
17

μm 

0.32mm 
 
8

μm 

0.90mm 
 
41

μm 

0.54mm 
 
17

μm 

Hardness 
 800HV 

Indentation 
diameter 
Depth of indentation 

0.35mm 
 
10

μm 

0.35mm 
 
10

μm 

0.35mm 
 
7

μm 

-- 
 
-- 

0.35mm 
 
10

μm 

Available types of impact devices 

DC:  test  hole  or 
hollow 

cylindrical 

DL: 

test 

slender 

narrow  groove  or 
hole 

D

+15: 

test 

groove 

or 

reentrant 
surface 

C: test small, light  
thin 

parts 

and 

surface 

of 

hardened layer 

G: 

test 

large, 

thick,  heavy  and 
rough 

surface 

steel 

E:

  test  super  high 

hardness material  

 
 
 
 
 
 
 
 
 
 

Summary of Contents for HMP

Page 1: ...Program 6 3 Fault Analysis Clearance 6 4 Note of Transport and Storage Conditions APPENDIX Table 1 Table 2 Table 3 Table 4 1 Overview 1 1 Advantages Wide measuring range Based on the principle of LEE...

Page 2: ...y Remarks Stand ard Config ura tion 1 Main unit 1 2 D type impact device 1 With cable 3 Standard test block 1 4 Cleaning brush I 1 5 Small support ring 1 6 Alkaline battery 2 AA size 7 Manual 1 8 Inst...

Page 3: ...To change the impact times in one group the key has to be pressed The impact times item will start flashing when the key is pressed Then the impact times value can be adjusted as desired by the keys a...

Page 4: ...the hardened layer should also conform to Table 3 Coupling Light weight samples must be firmly coupled with a heavy base plate Both coupled surfaces have to be flat and smooth there may not rest any...

Page 5: ...he type of impact device will be automatically detected while powering on This information will be displayed on the screen Users should take notice to this type After a few seconds the screen will exi...

Page 6: ...me during data logging 5 6 2 Clearing selected File Group It may be required to delete a file from the instrument s memory This procedure is outlined in following steps 1 The key has to be pressed to...

Page 7: ...Measurement data stored in memory of the gauge can be transferred to PC with the RS 232 port Detailed information of the communication software and its usage please refer to the software manual 5 13...

Page 8: ...tool steel HRC 20 4 67 1 19 8 68 2 20 7 68 2 22 6 70 2 HV 80 898 80 935 100 941 82 1009 Stainless steel HRB 46 5 101 7 HB 85 655 HV 85 802 Grey cast iron HRC HB 93 334 92 326 HV Nodular cast iron HRC...

Page 9: ...e surface Ra 1 6 m 1 6 m 0 4 m 6 3 m 1 6 m Min weight of sample Measure directly Need support firmly Need coupling tightly 5kg 2 5kg 0 05 2kg 5kg 2 5kg 0 05 2kg 1 5kg 0 5 1 5kg 0 02 0 5kg 15kg 5 15kg...

Page 10: ...sting cylindrical outside R25 R50 4 HZ11 13 For testing cylindrical inside R11 R13 5 HZ12 5 17 For testing cylindrical inside R12 5 R17 6 HZ16 5 30 For testing cylindrical inside R16 5 R30 7 K10 15 Fo...