To power on the instrument, the key 

 has to be pressed 

and held, meanwhile the key 

 is also pressed. Then the 

calibration screen is shown as on page 4. 
 
 
 
 
 
 
 
5 points have to be tested on the standard hardness block. 
After  that,  the  average  value  of  those  5  measured  values 

will  be  displayed.  The  keys   

  and 

  have  to  be 

pressed to change its nominal value. 
Finally the calibration has to be confirmed by pressing the 

  key.  To  cancel  the  calibration,  the 

  key  has  to  be 

pressed. Range of adjustment:  ±30 HL 
The  measurement  parameters,  including  the  material 
setting, the hardness scale and the impact direction, can’t 
be changed during calibration. 
 

3.2     Impact Device Selection 

For  the  selection  of  impact  devices,  please  refer  to 
APPENDIX Table 1 and Table 3. 
 

3.3     Preparation of the Sample Surface 

Preparation  of  the  sample  surface  should  conform  to  the 
relative requirements in APPENDIX Table 3. 

 

During  the  preparation  of  the  sample  surface, 
any  heating  or  cold  processing  on  this  surface 
should be avoided. 

 

Too  big  roughness  of  the  surface  being 
measured could cause errors. The surface of the 
sample  should  show  metallic  luster,  be  smooth 
and polished, without any stains of oil etc. 

 

Support  of  the  sample  is  not  necessary  for 
heavy  objects.  Medium-weight  samples  have  to 
be  placed  onto  a  smooth  and  stable  plane.  It 
must  be  placed  in  absolute  equability  and 
without any wobble. 

 

Curved  surface:  The  best  testing  surface,  of 
course,  is  flat.  When  the  curvature  radius  R  of 
the  surface  to  be  tested  is  smaller  than  30  mm 
(D,  DC,  D+15,  C,  E  and  DL  type  of  impact 
device)  and  smaller  than  50  mm  (G  type  of 
impact  device),  the  small  support  rings  should 
be chosen. 

 

The  sample  should  have  enough  thickness,  the 
minimum thickness of it should conform to  

        Table 3. 
*For  a  sample  with  a  hardened  layer  on  its  surface, 
the  depth  of the  hardened  layer  should also conform 
to Table 3. 
 
 
 

 

 

 

          
 
 
 
 

 

Coupling:  Light-weight  samples  must  be  firmly 
coupled  with  a  heavy  base  plate.  Both  coupled 
surfaces  have  to  be  flat  and smooth;  there may 
not  rest  any  redundant  coupling  agent.  The 
impact  direction  must  be  vertical  to  the  coupled 
surface. If the sample is  a big plate, long rod or 
bending  piece,  it  can  be  deformed  and  become 
unstable,  even  though  its  weight  and  thickness 
is  big  enough;  accordingly,  the  test  value  might 
not  be  accurate.  So  the  sample  should  be 
reinforced or supported at its back. 

 

Magnetism  of  the  sample  itself  should  be 
avoided.  

 

4.      Testing Program 
4.1    Start-up 

 

The plug of the impact device has to be inserted 
into  the  socket  of  the  impact  device  of  the 
instrument. 

 

The 

 key has to be pressed to power on. Now  

the instrument is in working mode. 

 

4.2     Testing 

-  The  release  button  on  the  upside  of  the  impact  device 
has  to  be  pressed  to  start  testing.  The  sample  and  the 
impact device as well as the operator are all required to be 
stable  now.  The  action  direction  should  pass  the  axis  of 
the impact device. 
-  Each  measurement  area  of  the  sample  usually  needs  3 
to 5 times of testing operations. The result data dispersion 
should not be more than the mean value ± 15 HL. 
- The distance between any two impact points or from the 
centre  of  any  impact  point  to  the  edge  of  the  sample 
should conform to the regulation of Table 4-1. 
-  If  an  accurate conversion  from  the  Leeb  hardness  value 
to  another  hardness  value  is  requested,  a contrastive test 
is  needed  to  get  conversion  relations  for  the  special 
material. An inspected qualified Leeb hardness tester and 
a corresponding hardness tester have to be used to test at 
the  same  sample  respectively.  For  each  hardness  value, 
each  homogeneous  measurement,  5  points  of  Leeb 
hardness value are needed.  They are in the surroundings 
of more than  three indentations  which  need conversion  of 
hardness.  The  Leeb  hardness  arithmetic  average  value 
and  the  corresponding  hardness  average  value  as 
correlative  value  respectively  are  used  to  establish  an 
individual  hardness  contrastive  curve.  The  contrastive 
curve  at  least  should  include  three  groups  of  correlative 
data. 

Summary of Contents for HMP

Page 1: ...Program 6 3 Fault Analysis Clearance 6 4 Note of Transport and Storage Conditions APPENDIX Table 1 Table 2 Table 3 Table 4 1 Overview 1 1 Advantages Wide measuring range Based on the principle of LEE...

Page 2: ...y Remarks Stand ard Config ura tion 1 Main unit 1 2 D type impact device 1 With cable 3 Standard test block 1 4 Cleaning brush I 1 5 Small support ring 1 6 Alkaline battery 2 AA size 7 Manual 1 8 Inst...

Page 3: ...To change the impact times in one group the key has to be pressed The impact times item will start flashing when the key is pressed Then the impact times value can be adjusted as desired by the keys a...

Page 4: ...the hardened layer should also conform to Table 3 Coupling Light weight samples must be firmly coupled with a heavy base plate Both coupled surfaces have to be flat and smooth there may not rest any...

Page 5: ...he type of impact device will be automatically detected while powering on This information will be displayed on the screen Users should take notice to this type After a few seconds the screen will exi...

Page 6: ...me during data logging 5 6 2 Clearing selected File Group It may be required to delete a file from the instrument s memory This procedure is outlined in following steps 1 The key has to be pressed to...

Page 7: ...Measurement data stored in memory of the gauge can be transferred to PC with the RS 232 port Detailed information of the communication software and its usage please refer to the software manual 5 13...

Page 8: ...tool steel HRC 20 4 67 1 19 8 68 2 20 7 68 2 22 6 70 2 HV 80 898 80 935 100 941 82 1009 Stainless steel HRB 46 5 101 7 HB 85 655 HV 85 802 Grey cast iron HRC HB 93 334 92 326 HV Nodular cast iron HRC...

Page 9: ...e surface Ra 1 6 m 1 6 m 0 4 m 6 3 m 1 6 m Min weight of sample Measure directly Need support firmly Need coupling tightly 5kg 2 5kg 0 05 2kg 5kg 2 5kg 0 05 2kg 1 5kg 0 5 1 5kg 0 02 0 5kg 15kg 5 15kg...

Page 10: ...sting cylindrical outside R25 R50 4 HZ11 13 For testing cylindrical inside R11 R13 5 HZ12 5 17 For testing cylindrical inside R12 5 R17 6 HZ16 5 30 For testing cylindrical inside R16 5 R30 7 K10 15 Fo...