Deep SUNA Manual
For SUNA running firmware version 2.4 or later
6. Use Scenarios
Setting Parameter
Value in Profiling Deployment
Input /
Output
Message Level
Warn
Message File Size
2
Output Frame
None.
Logging Frame
Full_ASCII
Logging Dark Frame
Output
Log File Type
Acquisition
Data
Acquisition
Operation Mode
Continuous
Operation Control
Samples
External Device
None
Countdown
15
Dark Averages
1
Light Averages
5
Dark Samples
1
Light Samples
60
Process
Temperature Compensation
Off
Salinity Fitting
On
Bromide Tracing
Off
Concentrations to fit
3
Dark Correction Method
SpecAverage
Absorbance Cutoff
1.3
Integration Time Adjustment
On
Fit Wavelength Low / High
217, 240
Table 21: Configuration parameters illustrating a profiling deployment.
6.2 Moored
6.2.1 Objectives and Considerations
In moored applications, power management, especially if running from battery, has to be
considered.
Moored applications typically have infrequent service intervals. As most environments
cause bio-fouling of the sensor, counter measures (e.g., a fouling guard) are necessary.
Copyright © 2013, Satlantic LP, All rights reserved
50