Deep SUNA Manual
For SUNA running firmware version 2.4 or later
4. Terminal Interface of the SUNA
External Device Minimum Interval
The external device minimum interval is set to
0.
An integrated wiper that could use this setting is only available for SUNA V2.
Table 13: Data acquisition configuration parameters
Parameter
Possible Values
Default Value
Short Name
Operation Mode
Continuous, Fixedtime, Periodic,
Polled, APF
Fixedtime
opermode
Operation Control
Duration, Samples
Samples
operctrl
Countdown [s]
0
–
3600
3
countdwn
Fixed Time Duration [s]
1
–
1000000
10
fixddura
Periodic Interval
1m, 2m, 5m, 6m, 10m, 15m, 20m,
30m, 1h, 2h, 3h, 4h, 6h, 8h, 12h, 24h
1h
perdival
Periodic Offset [s]
any value
0
perdoffs
Periodic Duration [s]
0
–
255
10
perddura
Periodic Samples
0
–
255
10
perdsmpl
Polled Timeout [s]
0
–
65535
10
polltout
APF Timeout [h]
0
–
100
10
apfatoff
Skip Sleep At Startup
On, Off
Off
skpsleep
Lamp Stabil. Time [ds]
0
–
255
5
stbltime
Lamp Switch-Off Temp. *
35
lamptoff
Spectrometer
Integration Period [ms]
5
–
60000
N/A
spintper
Dark Averages
1
–
200
1
drkavers
Light Averages
1
–
200
1
lgtavers
Dark Samples
1
–
65535
1
drksmpls
Light Samples
1
–
65535
10
lgtsmpls
Dark Duration [s]
1
–
65535
10
drkdurat
Light Duration [s]
1
–
65535
120
lgtdurat
External Device
None
None
exdevtyp
Ex. Dev. Pre-run time [s] 0
0
exdevpre
Ex. Dev. During Acq.
Off
Off
exdevrun
Ex. Dev. Min. Interval
0
0
exdvival
Copyright © 2013, Satlantic LP, All rights reserved
31