4.10 Alibi Memory
The instrument is equipped with an alibi memory for approx. 80000 items. If the number is exceeded, the
eldest items are overwritten. No license is required for using the alibi memory. At each print command a
dataset with a sequence number is filed. The dataset can comprise of up to 3 items (Gross, Net and Tare).
Each item consists of :
- Sequence number
1 - 999999
- Date
YYYY-MM-DD
- Time
hh:mm:ss
- Scale address
Always A
- Weight type
B = Gross, N = Net, T = Tare, PT = Preset tare
- Weight with unit
Separation of decimals is done by the decimal point
Depending on the configuration (Using front keys: oP 063) Gross- and/or Net- and/or Tare weight is filed,
see chapter 4.3.3 . Display of the datasets using VNC see chapter 4.4.3 .
The alibi memory can be completely erased, if the CAL switch is open, using front keys (AL 101).
With the Browser the datasets can be
exported as CSV file for later
evaluation as EXCEL file.
The CSV file uses the 'comma' as separator between the different elements of an item. Consequently in the
EXCEL program a conversion has to be activated to receive only one element per column.
The sequence number is preceded by an additional line number.
Example (Dataset as CSV-file):
Line,Seq.No.,Date,Time,WP,Type,Weight
55,13,2007-05-22,16:26:05,A,B,"12.450 kg"
56,13,2007-05-22,16:26:05,A,N,"10.400 kg"
57,13,2007-05-22,16:26:05,A,T,"2.050 kg"
After conversion in EXCEL:
Take care: EXCEL can import at maximum 65536 lines, if these are exceeded, the CSV-file has to be divided