Installing the Instrument and Plug-in Cards
X3 Operating Manual
18
Sartorius
3.3.2.1 Connecting a Remote Display / Remote Terminal
A PR 1627 remote display or a PR 1628 remote terminal can be connected to the built-in RS-232 interface
[Builtin RS232] or to the PR 5510/04 card.
Press -[Serial ports parameters]-[Remote display]-[Builtin RS232]-[Param] and select [Baudrate] 4800. The
baud rate now corresponds to the default settings of PR 1627 or PR 1628.
The following settings cannot be changed: [Bits] 7, [Parity] even and [Stopbits] 1.
If only one instrument is connected to a PR1627/8, [Mode] must be [single transmitter].
When connecting several instruments to a PR 1628 via an RS-232/RS-485 converter, selection for display on
PR 1628 is possible using addresses. For this purpose, select [multiple transmitters] as [Mode] in all
instruments, enter the instrument address under [Device Id] and the address of the subsequent instrument
under [Next Device Id].
3.3.2.2 Connecting a YDP12IS or YDP04IS Ticket Printer
The ticket printer YDP12IS-OCEUV or YDP04IS-OCEUV can be connected via [Builtin RS232] interface or the
RS-232 on card PR 5510/04.
If the printer is connected to the [Builtin RS232] port:
Press -[Serial ports parameter]-[Printer]-[Builtin RS232]-[Param] and configure the following settings
under [Protocol]: [RTS/CTS], [Baudrate]: 9600, [Bits]: 8, [Parity]: [none], [Stopbits]: 1 and [Output mode]:
[Raw].
The printer must be set to Line Mode (factory setting: Page Mode). Press the 'FEED' button to change
modes; please refer to the installation instructions delivered with the printer.
Summary of Contents for X3 PR 5410
Page 14: ...Process Indicator X3 Operating Manual 14 Sartorius 2 3 6 Overview of Accessories ...
Page 144: ......
Page 147: ......