Installing the Instrument and Plug-in Cards
X3 Operating Manual
24
Sartorius
3.4.4 PR 5510/04 Serial I/O
The plug-in card has two channels (1x RS-232 and 1x RS-422/485), which can be used simultaneously and
to a great extent independently. The RS-422/485 interface is electrically isolated. Up to 2 PR 5510/04 cards
(Slot 1 and 2) can be plugged in. The relevant interface parameters can be configured under -[Serial ports
parameter]; the DIL switch S101 must be set for RS-422/485 additionally.
Internal connection:
34-contact socket terminal strip
Number of channels:
1x RS-232, 1x RS-422/485
Type:
RS-232 full duplex
RS-422/485 full duplex (4-wire) *
RS-485 half duplex (2-wire) *
Transfer rate:
300 to 19k2 bit/s
RS-232C (V24)
signals:
Output: TXD, RTS, DTR
Input: RXD, CTS, DCD, RI
RS-422/485 signals:
TxA, RxA, TxB, RxB
Electrical isolation:
RS-232: no, RS-422 / 485: yes
Cable length:
Max. 15m with RS-232
Max. 1000 m with RS-422 / 485
Cable type:
Shielded twisted pair
(e.g., LifYCY 3x2x0,20),
1 conductor pair for GND.
External connection:
2 D-Sub 9-contact female
connectors
Accessories (delivered
with the unit):
2 D-Sub 9-pin mating plugs
(male)
incl. shielded housing
Dimensions: (LxWxH):
86 x 52 x 15 mm
Weight:
33 g
3.4.4.1 PR 5510/04 RS-232
The RS-232 interface is independent of the S101 switch settings.
It can be used only for point-to-point connection.
PR 5510/04 is provided as an equivalent to the [Builtin RS-232] interface in the RS-232 channel with
additional signals: DCD, DTR, RI.
The built-in and PR 5510/04 interfaces comply with the standard pin allocation; i.e., they are equivalent in
the following connecting diagrams. Thus only the RS-232 connections for the built-in interface are
described in this manual (see Chapter 3.3.2).