X3 Operating Manual
Installing the Instrument and Plug-in Cards
Sartorius
23
3.4.3 PR 5510/02 Serial I/O
The plug-in card contains two RS-232 channels (A and B), which can be used simultaneously and
independently. Max. 2 PR 5510/02 cards can be plugged in (Slot 1 ... 2). The relevant interface parameters
are adjustable in -[Serial ports parameter], no additional settings on the card are required.
Internal connection:
34-pole connector socket
Number of channels:
2
Type:
RS 232, full duplex
Transfer rate:
300 to 19k2 bits/sec
Parity:
No, odd, even
Data bits:
7 / 8 Bit
Signals RS 232:
Output: TX, RTS
Input: RX, CTS
Input signal level:
logic 1 (high) - 3 ... - 15 V
logic 0 (low) + 3 ... + 15 V
Output signal level:
logic 1 (high) - 5 ... - 15 V
logic 0 (low) + 5 ... + 15 V
Potential isolation:
No
Cable length:
max. 15m
Cable type:
twisted pairs, screened
(e.g. LifYCY 3x2x0,20),
1 conductor pair for GND.
External connection:
2x D-Sub 9-pole socket (female)
Accessories (delivered
with the unit):
2x connector counterpart D-Sub
9-pin (male)
incl. screening hoods
Dimensions: (LxWxH):
86 x 52 x 15 mm
Weight:
appr. 30 g
The RS-232 can only be used as point to point connection. A max. cable length of 10-15m must not be
exceeded.
The PR5510/02 and the 'Builtin' comply with the standard pin allocation, i.e. they are equal in the
connecting diagrams. Accordingly, the RS232 connections are described only for the builtin interface in this
Summary of Contents for X3 PR 5410
Page 14: ...Process Indicator X3 Operating Manual 14 Sartorius 2 3 6 Overview of Accessories ...
Page 144: ......